Rohde & Schwarz HZ115 Operation Manual (PDF)

The manual for the HZ 115 differential probe provides detailed instructions on setup, operation, and troubleshooting, including battery installation, connection to an oscilloscope, and safety precautions to ensure proper use and maintenance. It emphasizes the importance of using qualified personnel, adhering to voltage limits, and following safety standards to prevent damage or injury.

  • Brand: Rohde & Schwarz
  • Model: HZ115
  • Type: operation manual
  • File size: 144.9 KB

Manual Overview

The operation manual for the HZ 115 differential probe outlines essential procedures for safe and effective use, focusing on troubleshooting, diagnostics, and maintenance. The probe is designed for use with oscilloscopes, supporting a maximum input voltage of 1500 V and featuring a bandwidth of 30 MHz with attenuation settings of x100 and x1000. Troubleshooting involves checking the battery status, indicated by a flashing LED when voltage drops below 6 V, and ensuring the probe is properly connected to the oscilloscope via the BNC socket. Calibration is crucial, with the probe requiring a 1 MΩ input impedance for accurate readings. The manual emphasizes the importance of using high-quality, safety-compliant accessories and maintaining a clean, dry probe environment. Error diagnostics include automatic switch-off after 10 minutes of inactivity and a beep alert for overrange conditions. Board-level guidance involves careful handling during battery replacement to avoid damaging electronic components. The probe can operate on a 9 V battery or an external 12 V DC supply, with specific safety precautions to prevent electrical hazards.

Quick Issues Covered

  • Diagnose power-up failures with automatic switch-off and low battery alerts.
  • Identify error codes via beep for overrange conditions.
  • Use schematics for circuit tracing and component identification.

Overview

The HZ 115 is a differential probe designed for use with oscilloscopes. It supports a maximum input voltage of 1500 V and offers a bandwidth of 30 MHz. The probe features attenuation settings of x100 and x1000, and a common mode rejection ratio of over 50 dB at 1 MHz. It includes user-friendly features such as a microcontroller, automatic switch-off, overrange indication with a beep, and low battery indication. The probe can operate on a 9V battery or an external power supply. The manual covers safety precautions, items supplied, technical features, usage instructions, maintenance, and examples of applications.

Specifications

  • 1500 V input voltage max.
  • 30 MHz bandwidth
  • x 100 and x 1000 attenuation
  • > 50 dB CMRR (1MHz)
  • Automatic Switch-off
  • Low battery indication
  • Battery operation with external supply option
  • Input Differential input voltage max. + 1500V (DC+AC peak)
  • Common mode input voltage + 1500V (DC+AC peak)
  • Installation category of input CAT III
  • Operating temperature 0°C to +40°C
  • Storage temperature -10°C to +60°C
  • Relative humidity 80 % RH at 40°C
  • Input impedance 60MΩ ||1.5 pF
  • Accuracy after 1 min. + 3% (18°C – 30°C)
  • Frequency Bandwidth at x 100 20 MHz
  • Bandwidth at x 1000 30 MHz
  • Rise time at x 100 17.5 nsec
  • Rise time at x 1000 12 nsec
  • Output voltage max. + 1.5 V (1MΩ)
  • Output impedance 50Ω
  • Noise max. 1.5 mV
  • Common mode rejection ratio DC: 70 dB AC: >50 dB (at 1 MHz)
  • Battery 9V block
  • Low battery indication LED flashes at U batt < 6 V
  • Optional power supply 12V DC /35 mA
  • Operating time on one battery approx. 16 hours
  • Battery switch-off automatic, after 10 minutes
  • Dimensions Case (LxWxH) 157 x 60 x 26 mm
  • Weight 300 g with battery
  • Material ABS, internal screening
  • Output cable length 50 cm

Controls & Indicators

The front-panel controls and indicators of the HZ 115 differential probe include an ON button, an OFF button, an attenuation switch with positions for x100 and x1000, and a low battery indication LED that flashes when the battery voltage is below 6V.

Installation / Setup

The installation and setup of the HZ 115 differential probe involve several key steps and precautions. Before using the probe, it is essential to read the safety precautions thoroughly. The probe must be used by qualified personnel and properly earthed by connecting it to an oscilloscope via the BNC output socket, ensuring the oscilloscope is also earthed. The maximum input voltage should not exceed 1500 V, and the probe should not be exposed to humidity or aggressive environments. The probe is supplied with a 9 V battery, which must be inserted before use. If the battery indicator LED flashes, the battery needs replacement. An external power supply can be used, disconnecting the internal battery. When connecting the probe, ensure the oscilloscope’s input impedance is 1 MΩ, and use safety leads for measurements up to 1500 V.

Operation

Before using the differential probe, ensure to read the precautionary measures thoroughly. Insert the supplied 9V battery into the battery compartment and connect it using the clip. If the battery indicator LED flashes, replace the battery as it indicates low voltage. To replace the battery, remove the screws on the rear face, open the probe, replace the battery, and reassemble the probe carefully without damaging components.

For external power supply, connect the probe to a suitable power source using the side socket, ensuring the power supply voltage is between 9 – 16 V DC. The internal battery will disconnect, and the automatic switch-off feature will not function.

When connecting the probe, ensure the oscilloscope input impedance is 1 MQ. Connect the probe output to the oscilloscope input using the coaxial lead and BNC socket. Avoid extending the cable significantly due to high impedance. For oscilloscopes with a 50Q input, note that the probe output voltage is halved, and consumption may increase. Use safety leads for reliable measurements up to +1500V DC + peak AC. Ensure any accessories used conform to safety standards.

Safety & Warnings

The safety precautions for the HZ 115 differential probe for oscilloscopes include:

– The probe should only be used by qualified personnel.
– Proper earthing is essential: connect the probe to the oscilloscope using the BNC output socket to ensure it is earthed. Ensure the oscilloscope is also connected to earth according to regulations.
– Connect the probe to the oscilloscope before attaching the leads to the test item, and disconnect the leads before removing the probe from the oscilloscope.
– Use only high-quality accessories that meet safety standards.
– Do not open the probe while the input leads are connected.
– Do not exceed the maximum input voltage of 1500 V (DC + AC peak) or 1400 V when using equipment rated for 1000 V CAT III.
– Avoid exposing the probe to humidity, steam, or aggressive or explosive environments.
– Keep the case and connection leads clean and prevent moisture or liquid from entering the probe or its components.
– Do not use the probe if it appears to be malfunctioning or faulty.
– Ensure the mains power supply complies with current standards.
– Remove the battery if the probe will not be used for an extended period to prevent damage to the battery compartment.

Download PDF (144.9 KB)