The manual for the B&K Precision 889B Bench LCR/ESR Meter provides detailed instructions on setting up the device, performing measurements of inductors, capacitors, and resistors, and utilizing its built-in functions for DC/AC voltage/current measurements and diode/audible continuity checks. It also includes troubleshooting guidance, such as open/short compensation and selecting measurement modes, to ensure accurate and efficient operation in various applications like electronic engineering labs and service shops.
Manual Overview
The B&K Precision 889B Bench LCR/ESR Meter is a versatile instrument designed for measuring inductors, capacitors, and resistors with a basic accuracy of 0.1%. It supports both DC and AC voltage/current measurements and includes diode and audible continuity checks. The meter features auto and manual ranging, selectable test frequencies (100 Hz to 200 kHz), and test voltages (50 mVrms to 1 VDC). The dual display allows simultaneous measurements of primary and secondary parameters such as impedance, inductance, capacitance, resistance, phase angle, and equivalent series resistance. For troubleshooting, the manual provides a detailed flow for diagnosing issues using error codes and diagnostics. Calibration and adjustments are facilitated by open/short calibration procedures. Board-level guidance includes detailed instructions on connecting test leads and selecting measurement modes. The 889B is suitable for use in electronic engineering labs, production facilities, service shops, and educational institutions, offering functions comparable to more expensive LCR bridges and general-purpose digital multimeters.
Quick Issues Covered
- Diagnose LCR meter power-up failures and initialization issues.
- Error codes for common measurement faults and troubleshooting steps.
- Calibration and adjustment procedures for accurate component testing.
Overview
The B&K Precision Corp. 889B Synthesized In-Circuit LCR/ESR Meter is a high accuracy test instrument designed for measuring inductors, capacitors, and resistors with a basic accuracy of 0.1%. It also includes functions for DC/AC Voltage/Current measurements and Diode/Audible Continuity checks. The 889B can be used in various settings such as electronic engineering labs, production facilities, service shops, and schools. It features auto or manual ranging, multiple test frequencies and voltages, and a dual display for simultaneous measurements. Additionally, it can perform functions of a general-purpose Digital Multi-Meter with a basic accuracy of 0.4% in voltage and current measurements. The manual covers topics such as impedance parameters, specifications, accessories, physical description, measurement procedures, remote mode commands, applications, warranty, and safety precautions.
Specifications
- Voltage Measurements: True RMS, up to 600Vrms @ 40 ~ 1K Hz, up to 600V, Input Impedance: 1M-Ohm
- Current Measurements: True RMS, up to 2Arms @ 40 ~ 1K Hz, up to 2A, Current Shunt: 0.1 Ohm @ > 20mA; 10 Ohm @ < 20mA
- Diode/Audible Continuity Checks: Open Circuit Voltage: 5Vdc, Short Circuit Current: 2.5mA, Beep On: < 250, Beep Off: > 50 Q
- LCR Measurements: Test conditions Frequency 100Hz / 120Hz / 1KHz / 10KHz / 100KHz / 200KHz, Level 1Vrms / 0.25Vrms / 50mVrms / 1VDC (DCR only), Measurement Parameters: Z,Ls, Lp, Cs, Cp, DCR, ESR, D, Q and 9, Basic Accuracy: 0.1%, Dual Liquid Crystal Display, Auto Range or Range Hold, USB Interface Communication, Open/Short Calibration, Primary Parameters Display: Z, DCR, Ls, Lp, Cs, Cp, Second Parameter Display: s], ESR, D, Q, Combinations of Display: Serial Mode: Z-8,Cs—D, Cs—Q, Cs—ESR, Ls—D, LS—Q,_LS—ESR, Parallel Mode: Cp-—D, Cp-—Q,Lp-—D, Lp-Q