BK Precision 2630 Instruction Manual (3) (PDF)

The manual for Bk 3 covers the setup, operation, and troubleshooting of spectrum analyzers, including models 2165A, 2620A, 2625, and 2630, with detailed instructions on safety precautions, calibration, and frequency measurements. It also provides information on the specifications, optional accessories, and customer support for these devices.

  • Brand: BK Precision
  • Model: 2630
  • Type: manual
  • File size: 977.8 KB

Manual Overview

The manual for Bk 3 Spectrum Analyzers, covering models 2165A, 2620A, 2625, and 2630, provides detailed guidance on troubleshooting, error codes, calibration, and board-level adjustments. Troubleshooting is structured through a logical flow, emphasizing the identification and resolution of common issues. Error codes and diagnostics are integral, offering specific codes for various faults, aiding in quick identification of problems. Calibration procedures are outlined for both vertical and horizontal adjustments, ensuring accurate measurements and optimal performance. Board-level guidance includes detailed instructions for handling and replacing components, with a focus on maintaining the integrity of the device. Safety precautions are emphasized, particularly regarding electrical shock risks, with recommendations for using isolation transformers and adhering to safe work practices. The manual also specifies the importance of using appropriate tools and techniques to prevent damage to the equipment and ensure technician safety.

Quick Issues Covered

  • Diagnose power-up failures with step-by-step troubleshooting guide.
  • Resolve calibration issues for vertical and horizontal adjustments.
  • Use schematics for effective circuit tracing and diagnostics.

Overview

The manual covers the operation and safety guidelines for B+K Precision Spectrum Analyzers, specifically models 2165A, 2620A, 2625, and 2630. It includes sections on test instrument safety, specifications, optional accessories, general information, controls and indicators, calibration, and customer support. The manual also provides detailed specifications for frequency input, amplitude, and other technical aspects of the spectrum analyzers.

Specifications

  • Frequency Input
  • Frequency range: 0.15 MHz to 1050 MHz (Models 2625 and 2630), 0.15 MHz to 500 MHz (Models 2615A and 2620A)
  • Input impedance: 50Ω
  • Input connector: BNC
  • Input attenuator: 0 to 40 dB (4 x 10 dB steps)
  • Input attenuator accuracy: ±1 dB/10 dB step
  • Max. input level: +10 dBm, +25VDC (0 dB attenuation), +20 dBm (40 dB attenuation)
  • Center frequency display accuracy: ±100 kHz
  • Marker accuracy: ±(0.1% span + 100 kHz)
  • Frequency display resolution: 100 kHz
  • Frequency scanwidth: 100 kHz/div. to 100 MHz/div. (Models 2625 and 2630), 50 kHz/div. to 50 MHz/div. (Models 2615A and 2620A)
  • Frequency scanwidth accuracy: ±10%
  • Frequency stability: Drift: <150 kHz / hour
  • IF-Bandwidth (—3 dB): Resolution: 400 kHz and 20 kHz (Models 2625 and 2630), 250 kHz and 20 kHz (Models 2615A and 2620A)
  • Video-Filter on: 4 kHz
  • Sweep rate: 43 Hz
  • Amplitude
  • Amplitude range: —100 dBm to +13 dBm
  • Screen display range: 80 dB (10 dB/ div.)
  • Reference level: —27 dBm to +13 dBm (in 10 dB steps)
  • Reference level accuracy: ±2 dB
  • Average noise level: —99 dBm (20 kHz BW)
  • Tracking Generator (Models 2620A and 2630 only)
  • Output level range: —50 dBm to +1 dBm (in 10 dB steps and variable)
  • Output attenuator: 0 to 40 dB (4 x 10 dB steps)
  • Output attenuator accuracy: ±1 dB
  • Output impedance: 50Ω (BNC)
  • Frequency range: 0.15 MHz to 1050 MHz (Model 2630), 0.1 MHz to 500 MHz (Model 2620A)
  • General
  • Display: CRT, 6 inch, 8 x 10 div. internal graticule
  • Trace rotation: Adjustable on front panel
  • Output Probe Power: 6V
  • Line voltage: 115 /230V ±10%, 50-60Hz
  • Power consumption: approx. 27W
  • Operating ambient temperature: +10°C to +40°C
  • Protective system: Safety Class I (IEC 1010-1)
  • Weight: Approx. 15.4 lb. (6 kg) (Models 2625 and 2630)

Safety & Warnings

Normal use of test equipment exposes users to electrical shock hazards. A current of 10 milliamps passing through the heart can be lethal, and voltages as low as 35 volts can produce a lethal current under certain conditions. To reduce risk, follow these precautions:

1. Avoid unnecessary exposure to high voltage. Remove housings and covers only when necessary and turn off equipment when making test connections in high-voltage circuits. Discharge high-voltage capacitors after removing power.
2. Familiarize yourself with the equipment being tested and the location of its high voltage points. Be aware that high voltage may appear unexpectedly in defective equipment.
3. Use insulated floor material or a large, insulated floor mat to stand on, and an insulated work surface for equipment. Ensure these surfaces are not damp or wet.
4. Use the “one hand in the pocket” technique when handling an instrument probe to avoid creating a ground return path.
5. Remember that ac line voltage is usually present on some power input circuits whenever equipment is connected to an ac outlet, even if turned off.
6. Some equipment with a two-wire ac power cord may have a “hot chassis.” Use an isolation transformer when testing such equipment to prevent shock hazards and damage.
7. Use a 3-wire outlet for equipment with a 3-wire ac power plug to keep exposed elements at earth ground.
8. B+K Precision products are not authorized for use in direct contact with the human body or as critical components in life support devices.
9. Never work alone. Ensure someone is nearby to provide aid if necessary, and training in CPR is highly recommended.