BK Precision 2630 Instruction Manual (2) (PDF)

The manual for BK Precision Spectrum Analyzers models 2165A, 2620A, 2625, and 2630 provides detailed instructions on setup, operation, and troubleshooting, including safety precautions for handling high voltage equipment. It covers specifications, calibration procedures, and maintenance tips to ensure accurate frequency and amplitude measurements.

  • Brand: BK Precision
  • Model: 2630
  • Type: manual
  • File size: 1.1 MB

Manual Overview

The manual for Bk Precision Spectrum Analyzers, models 2165A, 2620A, 2625, and 2630, provides comprehensive guidance on troubleshooting, error diagnostics, calibration, and board-level adjustments. Troubleshooting involves a systematic flow to identify and resolve issues, with emphasis on understanding error codes and diagnostics to pinpoint faults. Calibration procedures are detailed for both vertical and horizontal adjustments to ensure accurate readings, with specific instructions for each model. The manual includes board-level guidance, highlighting the importance of familiarizing oneself with the equipment’s high voltage points and using insulated materials to prevent electrical shock. Safety precautions are emphasized, such as using isolation transformers for “hot chassis” equipment and adhering to proper grounding techniques. The manual also specifies the frequency range, input impedance, and other technical specifications necessary for effective operation and maintenance of the spectrum analyzers.

Quick Issues Covered

  • Diagnose power-up failures with detailed troubleshooting steps.
  • Identify and resolve error codes specific to Model 2620A.
  • Use schematics for effective circuit tracing and repairs.

Overview

The source provides information about BK Precision spectrum analyzers, specifically models 2165A, 2620A, 2625, and 2630. It includes safety precautions for using test instruments, specifications for frequency input and amplitude, and general operating conditions. The manual covers topics such as introduction to spectrum analysis, controls and indicators, calibration, and customer support.

Specifications

  • Frequency Input
  • Frequency range: 0.15 MHz to 1050 MHz (—3 dB) (Models 2625 and 2630)
  • Frequency range: 0.15 MHz to 500 MHz (-3 dB) (Models 2615A and 2620A)
  • Input impedance: 50Ω
  • Input connector: BNC
  • Input attenuator: 0 to 40 dB (4 x 10 dB steps)
  • Input attenuator accuracy: ±1 dB/10 dB step
  • Max. input level: +10 dBm, +25VDC (0 dB attenuation)
  • Max. input level: +20 dBm (40 dB attenuation)
  • Center frequency display accuracy: ±100 kHz
  • Marker accuracy: ±(0.1% span + 100 kHz)
  • Frequency display resolution: 100 kHz
  • Frequency scanwidth: 100 kHz/div. to 100 MHz/div. (Models 2625 and 2630)
  • Frequency scanwidth: 50 kHz/div. to 50 MHz/div. (Models 2615A and 2620A)
  • Frequency scanwidth accuracy: ±10%
  • Frequency stability: Drift: <150 kHz / hour
  • IF-Bandwidth (—3 dB): Resolution: 400 kHz and 20 kHz (Models 2625 and 2630)
  • IF-Bandwidth (—3 dB): Resolution: 250 kHz and 20 kHz (Models 2615A and 2620A)
  • Video-Filter on: 4 kHz
  • Sweep rate: 43 Hz
  • Amplitude
  • Amplitude range: —100 dBm to +13 dBm
  • Screen display range: 80 dB (10 dB/ div.)
  • Reference level: —27 dBm to +13 dBm (in 10 dB steps)
  • Reference level accuracy: ±2 dB
  • Average noise level: —99 dBm (20 kHz BW) (Models 2625 and 2630)
  • Tracking Generator (Models 2620A and 2630 only)
  • Output level range: —50 dBm to +1 dBm (in 10 dB steps and variable)
  • Output attenuator: 0 to 40 dB (4 x 10 dB steps)
  • Output attenuator accuracy: ±1 dB
  • Output impedance: 50Ω (BNC)
  • Frequency range: 0.15 MHz to 1050 MHz (Model 2630)
  • Frequency range: 0.1 MHz to 500 MHz (Model 2620A)
  • Frequency response: ±1.5 dB
  • Radio Frequency Interference (RFI): <20 dBc
  • General
  • Display: CRT, 6 inch, 8 x 10 div. internal graticule
  • Trace rotation: Adjustable on front panel
  • Output Probe Power: 6V
  • Line voltage: 115 /230V ±10%, 50-60Hz
  • Power consumption: approx. 27W
  • Operating ambient temperature: +10°C to +40°C
  • Protective system: Safety Class I (IEC 1010-1)
  • Weight: Approx. 15.4 lb. (6 kg) (Models 2625 and 2630)
  • Weight: Approx. 13.2

Safety & Warnings

Normal use of test equipment exposes users to electrical shock hazards, as testing often involves exposed voltage. A current of 10 milliamps passing through the heart can be lethal, and voltages as low as 35 volts DC or AC RMS should be considered dangerous. Higher voltages increase the risk of lethal current. To reduce risk, follow these safety precautions:

1. Some equipment with a two-wire AC power cord may have a “hot chassis.” When servicing, use an isolation transformer to prevent shock hazards and damage to test instruments. Treat all two-wire AC equipment as “hot-chassis” unless confirmed otherwise.

2. Use a 3-wire outlet for test instruments or equipment with a 3-wire AC power plug to ensure safety by keeping exposed elements at earth ground.

3. B+K Precision products are not authorized for direct contact with the human body or as critical components in life support devices.

4. Never work alone. Have someone nearby to provide aid if necessary, and training in CPR is highly recommended.

5. Avoid exposing high voltage unnecessarily. Remove housings and covers only when needed, turn off equipment when making test connections, and discharge high-voltage capacitors after removing power.

6. Familiarize yourself with the equipment and high voltage points, but be aware that high voltage may appear unexpectedly in defective equipment.

7. Use insulated floor material or a large, insulated floor mat, and ensure work surfaces are not damp or wet.

8. Use the “one hand in the pocket” technique while handling instrument probes to avoid creating a ground return path.

9. Remember that AC line voltage is usually present on some power input circuits whenever equipment is connected to an AC outlet, even if turned off.

Download PDF (1.1 MB)