Heathkit EUW-20A Servo Recorder Schematics (PDF)

The Heathkit schematics manual provides detailed instructions for setting up and operating the device, including specifications for chart paper, input circuits, and power requirements. It also offers troubleshooting guidance by outlining the schematic diagrams and component values necessary for diagnosing and resolving issues.

  • Brand: Heathkit
  • Model: EUW-20A
  • Type: manual
  • File size: 163.8 KB

Manual Overview

The Heathkit SCHEMATICS manual provides detailed guidance for troubleshooting, diagnostics, and calibration of the Heath EUW-20A Servo-Recorder. Troubleshooting flow involves checking the self-balancing potentiometric input circuit, which maintains an infinite resistance at null and approximately 500 KΩ off null. Error diagnostics can be conducted by verifying the overall error, which should be less than 1% of full scale from 10 to 250 mV, and ensuring the dead zone remains below 0.5% of full scale. Calibration involves adjusting the chart span across five fixed ranges with additional vernier adjustments, allowing for precise span settings between 250 and 3.3 mV. Board-level guidance includes checking the power requirements, ensuring the 1 ampere slow-blow fuse is intact, and verifying connections through accessory sockets. The manual specifies the use of a mercury cell reference source with a life of approximately 300 hours. For accurate voltage readings, use an 11 MΩ input VTVM with the circuit connected as shown, ensuring the system is balanced and servo power is on.

Quick Issues Covered

  • Diagnose power-up failures with fuse and power switch checks.
  • Adjust chart span using vernier for precise calibration.
  • Trace circuit issues using detailed schematic for model EUW-20A.

Specifications

  • Grid Width: 10"
  • Length: 120 foot roll
  • Markings: 0-100, right to left
  • Chart Speed: 2 inches per minute; additional motors for special chart speeds are available
  • Chart Span: Five fixed ranges: 10, 25, 50, 100, and 250 mv; plus a vernier adjustment between each fixed range
  • Balancing Time: 0.1 second per inch, 1 second full scale (10"). (50 cps, 1.5 seconds full scale)
  • Input Circuit: Self-balancing potentiometric on all ranges
  • Input Resistance: Essentially infinite at null, Approximately 500 KQ off null
  • Overall Error: Less than 1% of full scale from 10 to 250 mv
  • Dead Zone: Less than 0.5% of full scale from 10 to 250 mv
  • Linearity: Less than 0.5% full scale
  • Maximum Recommended Source Resistance: 50 KQ
  • Reference Source: Mercury cell
  • Reference Cell Life: 300 hours (approximately)
  • Accessory Sockets: 10 watts may be drawn from each socket at 120 V AC. 10 watt total may be drawn at 240 V AC. 6.3 V AC at 50 ma may be drawn from the accessory 11-pin socket
  • Power Requirements: 120/240 volts, 60 cps; 45 watts, (120/240 volts, 50 cps; 50 watts)
  • Fuse: 1 ampere slow-blow
  • Dimensions: 13-3/4" wide x 8-3/4" high x 13-3/8" deep
  • Net Weight: 17 lbs
  • Shipping Weight: 20 lbs

Download PDF (163.8 KB)

Leave a Comment