The Valhalla Scientific 2734A Operation and Maintenance Manual provides detailed instructions for unpacking, installation, initial adjustments, and operation of the direct voltage reference bank, including both manual and remote operation procedures. It also includes sections on troubleshooting, maintenance, calibration, and performance verification to ensure the device meets its specifications, with additional guidance on safety precautions and environmental considerations.
Overview
The 2734A Direct Voltage Reference Bank is a precision instrument certified by Valhalla Scientific, Inc. to meet its published specifications upon shipment. The manual covers various aspects of the product, including unpacking and installation, specifications, available options, front and rear panel controls, manual and remote operation, calibration, maintenance and troubleshooting, theory of operation, performance verification, useful hints, manual change information, schematic and assembly diagrams, and parts lists.
Specifications
- General specifications of the 2734A Direct Voltage Reference Bank are listed.
- Stability:
- – 1.0V: 1.0ppm (30 days), 2.0ppm (90 days), 3.0ppm (180 days), 5.0ppm (1 year)
- – 1.018V: 1.0ppm (30 days), 2.0ppm (90 days), 3.0ppm (180 days), 5.0ppm (1 year)
- – VZ: 0.4ppm (30 days), 0.8ppm (90 days), 1.6ppm (180 days), 3.0ppm (1 year)
- – 10V: 0.5ppm (30 days), 1.3ppm (90 days), 2.5ppm (180 days), 4.0ppm (1 year)
- Settability:
- – 1.0V: 10ppm total adjustment range, 0.02ppm resolution
- – 1.018V: 10ppm total adjustment range, 0.02ppm resolution
- – VZ (4 refs): 70ppm total adjustment range, 0.14ppm resolution
- – VZ (1 ref): 280ppm total adjustment range, 0.56ppm resolution
- – 10V: 5ppm total adjustment range, 0.01ppm resolution
- Temperature Coefficients:
- – 1.0V: 0.1ppm/C
- – 1.018V: 0.1ppm/C
- – VZ: 0.02ppm/C
- – 10V: 0.05ppm/C
- Temperature Shock:
- – Non-accumulative change in output voltage no greater than 0.01ppm/C change
- Output Drive and Resistance:
- – 1.0V: 857.14 ohm +/- 0.005%
- – 1.018V: 869.95 ohm +/- 0.005%
- – VZ: 3mA, 0.05 ohm maximum
- – 10V: 12mA, 0.005 ohm maximum
- Supply Regulation:
- – Maximum change in output voltage less than 0.05ppm following a 10 second delay
Installation / Setup
The installation/setup highlights for the 2734A Direct Voltage Reference Bank include the following:
1. Unpacking: If the shipping carton is damaged, ensure the carrier’s agent is present during unpacking. If damage is found, notify the carrier’s agent for repair authorization before returning the unit to Valhalla Scientific.
2. Initial Adjustments: Before operation, select the correct local power source voltage and verify the correct fuse is fitted. The supply voltages and corresponding fuses are:
– 105 to 128VAC 50/60Hz with a 1 Amp Slo Blo Fuse
– 210 to 256VAC 50/60Hz with a 0.5 Amp Slo Blo Fuse
3. Bench Use: The unit is ready for bench use without special instructions. Ensure the power cord has a three-terminal connector.
4. Rack Mounting: Optional brackets are available for 19″ rack mounting. The unit should be supported on both sides to prevent movement during transport. Allow sufficient airflow by placing 1.75″ high blank panels above and below the unit in the rack. Avoid ambient temperatures above 50°C during operation or 70°C when not in use.
5. Safety Precautions: Use a three-contact power connector with a continuous ground connection to ensure safety.