Sencore FC221 Field Calibrator for Sencore Z Meters Operator Manual (PDF)

The manual for the Sencore FC221 Field Calibrator provides instructions for setting up and operating the device to test the calibration accuracy of Sencore LC53 or CA55 analyzers, emphasizing the importance of recalibrating the FC221 annually to maintain accuracy. It also details troubleshooting steps, such as checking if readings fall within specified tolerances to determine if recalibration of the internal controls is necessary.

  • Brand: Sencore
  • Model: FC221
  • Type: manual
  • File size: 2.6 MB

Manual Overview

The Sencore FC221 Field Calibrator is designed to test the calibration accuracy of Sencore LC53 and CA55 analyzers. It uses low-drift coils and capacitors to provide a convenient field check of these instruments. The FC221 helps determine if the LC53 or CA55 is within tolerance, thereby reducing unnecessary recalibration. If readings fall outside the specified tolerance, recalibration of the internal controls is necessary. The FC221 does not cover all values needed for complete recalibration but tests most internal calibration controls. Calibration errors require additional standard components for correction. The FC221 should be recalibrated annually to maintain accuracy. When evaluating results, consider the combined tolerance of the FC221 and the LC53 or CA55. For instance, a reading between 0.459 and 0.480 uF is acceptable for a marked value of 0.470 uF. If readings are outside this range, either the analyzer or the calibrator may be out of tolerance. Always recalibrate the FC221 when servicing the LC53 or CA55 to ensure accurate field tests.

Quick Issues Covered

  • Diagnose calibration errors in LC53 or CA55 using FC221 values.
  • Identify out-of-tolerance readings requiring recalibration of internal controls.
  • Ensure FC221 recalibration annually to maintain accuracy standards.

Overview

The FC221 Field Calibrator is designed to test the calibration accuracy of the Sencore LC53 Meter Capacitor/Inductor Analyzer or the CA55 Capacitor Analyzer. It provides a set of low-drift coils and capacitors for convenient field checks of these instruments. The FC221 helps determine if the LC53 or CA55 is within tolerance, indicating whether recalibration is necessary. The calibrator needs annual testing to maintain its specifications and does not provide all values needed for complete recalibration but tests most internal calibration controls.

Specifications

  • Operating temperature range: 65 to 75° F (18 to 24° C).
  • Approximate Value Accuracy*
  • 8mH 0.75%
  • 560 uH 0.75%
  • 50 uH 0.75%
  • 5600 pF 0.75%
  • 0.47 uF 0.75%
  • 700 uF 1.5%
  • Calibration interval: One year maximum.
  • Size: 442” X 334” X112”’ (11.43 X 9.53 X3.81 cm.)
  • Weight: 3.5 oz. (98 g)

Safety & Warnings

WARNING: Do not check the 700 uF capacitor for leakage at a voltage greater than 6 volts, as permanent damage to the capacitor will result.