Rohde & Schwarz RTM-ZP10 Operation Manual (PDF)

The manual for the R&S RTM-ZP10 500 MHz Passive Voltage Probe provides detailed instructions on setup, operation, and troubleshooting, including safety precautions, electrical specifications, and adjustment procedures for low and high-frequency compensation. It also covers compliance with safety standards and environmental specifications for proper usage and maintenance.

  • Brand: Rohde & Schwarz
  • Model: RTM-ZP10
  • Type: operation manual
  • File size: 718.8 KB

Manual Overview

The operation manual for the R&S® RTM-ZP10 500 MHz Passive Voltage Probe provides detailed guidance on troubleshooting, error diagnostics, calibration, and board-level adjustments. For troubleshooting, the manual emphasizes ensuring proper grounding and connection sequences to prevent damage or inaccurate readings. Error diagnostics involve checking for compliance with specified voltage and frequency limits, as exceeding these can lead to probe failure. Calibration procedures include low-frequency (LF) and high-frequency (HF) compensation adjustments. LF compensation is crucial when initially connecting the probe to an oscilloscope, ensuring the probe cable capacitance matches the oscilloscope input capacitance, which is vital for maintaining amplitude accuracy across the frequency range. HF compensation adjustments are necessary for optimizing the probe’s response at higher frequencies. The manual also highlights the importance of adhering to safety standards and environmental conditions, such as operating temperature and humidity limits, to ensure reliable probe performance.

Quick Issues Covered

  • LF compensation adjusts probe capacitance for accurate amplitude readings.
  • Diagnose power-up failures with voltage derating and input impedance specs.
  • Use schematics for circuit tracing and troubleshooting probe connection issues.

Overview

The document provides information about the R&S® RTM-ZP10, a 500 MHz passive voltage probe manufactured by Rohde & Schwarz. It includes safety standards compliance, electrical and mechanical specifications, and environmental conditions for operation. The manual also outlines safety precautions, measurement categories, pollution degrees, and adjustment procedures for the probe.

Specifications

  • Attenuation Ratio: 10:1 +2 % at DC
  • Voltage Coefficient: 0.0025 %/V (typical)
  • System Bandwidth: 500 MHz (-3 dB)
  • Probe Risetime: 700 ps (10 % – 90 %) (typical)
  • Input Resistance (System): 10 MQ +1%
  • Input Capacitance (System): 9.5 pF (typical)
  • Compensation Range: 5 pF – 20 pF (typical)
  • Input Coupling of the Measuring Instrument: 1MQAC/DC
  • Maximum Rated Input Voltage:
  • Measurement Category I: 400 V root mean square, 1250 V transient overvoltage
  • Measurement Category II: 300 V root mean square CAT II
  • Pollution Degree: 2
  • Weight (Probe only): 48 9
  • Cable Length: 1.3m
  • Probe Tip Diameter: 2.5mm
  • Altitude operating: up to 2000 m
  • Altitude non-operating: up to 15000 m
  • Temperature Range operating: 0° C to +50° C
  • Temperature Range non-operating: -40° C to +71° C
  • Maximum Relative Humidity — operating: 80 % relative humidity for temperatures up to +31° C, decreasing linearly to 40 % at +50° C

Safety & Warnings

To avoid personal injury and to prevent fire or damage to the product or connected products, the following safety precautions should be reviewed and complied with:

– Only qualified personnel should use the probe assembly.
– Use only grounded instruments.
– Do not connect the probe ground lead to a potential other than earth ground. Ensure the probe and measurement instrument are properly grounded.
– Connect the probe output to the measurement instrument and the ground lead to earth ground before connecting the probe to the circuit under test. Disconnect the probe input and ground lead from the circuit under test before disconnecting from the measurement instrument.
– Observe probe ratings and do not exceed the maximum ratings of the probe.
– Comply with the voltage versus frequency derating curve.
– Keep away from live circuits and avoid open circuitry. Do not touch connections or components when power is present.
– Do not operate with suspected failures and refer to qualified service personnel.
– The product is for indoor use only and should not be operated in wet/damp environments. Keep surfaces dry and clean.
– Do not operate the product in an explosive atmosphere.