Rohde & Schwarz HZO30 Operation Manual (PDF)

The manual for the HZ030 active probe covers its setup, operation, and troubleshooting, emphasizing the importance of proper grounding and connection to measurement devices with a 50Ω input. It provides detailed specifications, safety instructions, and guidelines for handling and cleaning the probe to ensure optimal performance and prevent damage.

  • Brand: Rohde & Schwarz
  • Model: HZO30
  • Type: operation manual
  • File size: 745.5 KB

Manual Overview

The HZ030 active probe by HAMEG Instruments is designed for precise measurements across a wide frequency range, offering high input impedance and low capacitance. For troubleshooting, ensure the probe is connected to a grounded measuring device with a 50Ω input coupling. Calibration involves maintaining a maximum possible distance between the power cable and the probe tip. The probe is protected against static overvoltage, but applying voltages beyond specified limits can damage the amplifier at the probe tip. Error diagnostics focus on ensuring proper grounding and connection before measurement. For board-level adjustments, the probe can be used with any measuring instrument featuring a standard BNC connector. The probe’s input impedance decreases with increasing signal frequency, so adherence to maximum input voltage and dynamic measuring range is crucial to avoid linearity errors and damage. Regular maintenance includes careful handling of the probe tip and cable, and cleaning with a soft cloth dampened with distilled water or isopropyl alcohol.

Quick Issues Covered

  • Diagnose linearity errors by monitoring input voltage and range limits.
  • Identify power-up failures with proper grounding and connection checks.
  • Adjust input coupling to 50Ω for optimal active probe operation.

Overview

The HZ030 active probe by HAMEG Instruments, a Rohde & Schwarz company, is designed for inline probing systems, enabling simultaneous contact with adjacent pins in a 2.54 mm grid. It features a T-profile housing for easy attachment to pin headers, allowing for multiple parallel measurements. The probe is system-independent and connects to any measuring instrument with a 50Ω input via a standard BNC connector. It has an input resistance of 1MΩ and an input capacitance of 0.9 pF, making it suitable for measurements across all frequency ranges. The HZ030 offers high input impedance up to the GHz range, ideal for daily measurement tasks. The manual covers safety instructions, handling, cleaning, warranty, and repair information.

Specifications

  • Teilungsfaktor: 10:1 +0,5 % bei DC
  • Dynamischer Messbereich: +8 V
  • System Bandbreite: 600 MHz (-3 dB)
  • Eigenbandbreite: 1 GHz
  • Maximale bemessene Eingangsspannung: 20V
  • Eingangs-Widerstand: >1MQ +1%
  • Eingangs-Kapazitat: 0,9 pF [typisch]
  • Eingangskopplung des Messinstrumentes: 500 AC / DC
  • Uberschwingen: 6% [typisch]
  • Gewicht (nur Tastkopf): 96 g
  • Kabellange: 13m
  • Einsatzhohe in Betrieb: bis zu 2000 m
  • Einsatzhohe auBer Betrieb: bis zu 13000 m
  • Temperaturbereich in Betrieb: 0° C bis +50° C
  • Temperaturbereich auBer Betrieb: -40°C bis +70° C
  • Maximale relative Luftfeuchtigkeit in Betrieb: 80 % bei Temperaturen bis zu 30° C, linear fallend bis 40 % bei 50° C

Safety & Warnings

To avoid personal injury and prevent damage to the product and connected devices, follow these safety precautions:

– Use this accessory only in enclosed rooms.
– Never use this accessory in damp environments or under the influence of fumes. Keep the product dry and clean.
– Do not use the product in explosive environments.
– This accessory should only be used by technically qualified personnel.
– Use only grounded measuring devices.
– Connect this probe only to grounded measuring devices. Always ensure that the probe and the measuring device are properly grounded.
– Connect and disconnect properly. Connect the probe’s connector to the measuring device and connect the ground lead to earth before contacting the probe tip with the measurement point. Remove the probe tip and ground lead from the measurement point before disconnecting the probe from the measuring device.
– Do not apply any potential to the probe that exceeds the maximum rated values. Ensure compliance with the voltage reduction curve described on page 8.
– Stay away from dangerous circuits.
– Avoid working on open and unprotected circuits. Stay away from connections or components that pose a risk of contact.
– Use only flawless measuring accessories.
– Have this product serviced only by technically qualified personnel.

Download PDF (745.5 KB)