Rohde & Schwarz FSP-B30 B31 B32 Operation Manual (PDF)

The manual provides detailed instructions for the installation, operation, and troubleshooting of the R&S FSP-B30 DC Power Supply and R&S FSP-B31/32 Battery Pack, including safety guidelines and technical specifications. It covers the setup of the DC power supply and battery pack, their operation in various conditions, and procedures for charging, removing, and inserting batteries.

  • Brand: Rohde & Schwarz
  • Model: FSP-B30
  • Type: operation manual
  • File size: 1.1 MB

Manual Overview

The operation manual for the R&S FSP-B30 DC Power Supply and R&S FSP-B31/B32 Battery Pack primarily focuses on providing guidance for installation, operation, and troubleshooting of these components when used with R&S FSP Spectrum Analyzers and R&S ESPI Test Receivers. The manual details the installation process for both the DC power supply and battery pack, emphasizing the importance of adhering to safety instructions and ensuring proper connections. Troubleshooting involves checking connections and ensuring the correct voltage and current levels are maintained. Error codes and diagnostics are not explicitly detailed, but users are advised to refer to the manual’s technical specifications for operating conditions. Calibration and adjustments are not covered in depth, suggesting these components are designed for straightforward installation and use. Board-level guidance is not provided, indicating that the manual is intended for users with basic technical knowledge rather than for in-depth repairs or modifications. The manual also includes safety instructions for handling batteries, highlighting the need for proper disposal and adherence to safety regulations.

Quick Issues Covered

  • Diagnose DC power supply issues with R&S FSP-B30.
  • Resolve battery pack charging errors in R&S FSP-B31/32.
  • Trace circuit schematics for power-up failures.

Overview

The product is a DC Power Supply and Battery Pack designed for the R&S FSP Spectrum Analyzer and R&S ESPI Test Receiver. The manual covers technical information, safety instructions, installation, operation, and specifications for the R&S FSP-B30 DC Power Supply and R&S FSP-B31/B32 Battery Pack.

Specifications

  • R&S FSP-B30 DC Power Supply
  • Input voltage range 10 VDC to 28 VDC
  • Output voltage 120 VDC to 360 VDC / 300 W
  • Temperature range 0 °C to +50 °C
  • Storage temperature range -40 °C to +70 °C
  • Dimensions 145 mm x 154 mm x 65 mm
  • Weight 0,6 kg
  • R&S FSP-B31/-B32 Battery Pack
  • Input external battery 10 VDC to 28 VDC
  • Input Power supply (battery charge) 24 VDC / max. 3A
  • Output voltage 10 VDC to 28 VDC /10A
  • Battery operation 13.2 VDC / 200 Wh
  • Charge Time 5 h at 25 °C
  • Temperature range (discharge) 0 °C to +50 °C
  • Temperature range (charge) +10 °C to +40 °C
  • Storage temperature range (< 1 year) -20 °C to +35 °C
  • Storage temperature range (< 1 month) -20 °C to +55 °C
  • Dimensions 400 mm x 184 mm x 42 mm
  • Weight 3.7 kg
  • AC adapter (FSP-B31 only)
  • Input voltage range 100 VAC to 240 VAC + 10%
  • Input frequency range 50 Hz to 60 Hz +5%
  • Typical input power 140 VA
  • Output voltage 24V
  • Output current 3A
  • Temperature range 0 °C to +50 °C
  • Storage temperature range -20 °C to +70 °C
  • Dimensions 132 mm x 58 mm x 30 mm
  • Weight 0,3 kg

Operation

The source does not provide specific information on the operation, modes, or basic usage steps for the R&S FSP-B30 or R&S FSP-B31/32.

Safety & Warnings

This unit has been designed and tested in accordance with the EC Certificate of Conformity and has left the manufacturer’s plant in a condition fully complying with safety standards. To maintain this condition and to ensure safe operation, the user must observe all instructions and warnings given in this operating manual.

The unit may be used only in the operating conditions and positions specified by the manufacturer. Unless otherwise agreed, the following applies to R&S products: IP degree of protection 2X, pollution severity 2, overvoltage category 2, only for indoor use, altitude max. 2000 m. The unit may be operated only from supply networks fused with max. 16 A.

Unless specified otherwise in the data sheet, a tolerance of +10% shall apply to the nominal voltage and of +5% to the nominal frequency. For measurements in circuits with voltages Ving > 30V, suitable measures should be taken to avoid any hazards. In such cases it must be ensured that the power plug is easily reachable and accessible at all times (length of connecting cable approx. 2 m).

Units of protection class I with disconnectible AC supply cable and appliance connector may be operated only from a power socket with earthing contact and with the PE conductor connected. It is not permissible to interrupt the PE conductor intentionally, neither in the incoming cable nor on the unit itself as this may cause the unit to become electrically hazardous.

Any extension lines or multiple socket outlets used must be checked for compliance with relevant safety standards at regular intervals. If the unit has no power switch for disconnection from the AC supply, the plug of the connecting cable is regarded as the disconnecting device. Functional or electronic switches are not suitable for providing disconnection from the AC supply.

If the unit is to be permanently wired, the PE terminal of the unit must first be connected.