Combined operating and service manual for the Hewlett-Packard 333A and 334A distortion analyzers (revision of October 1975). The 334A adds an AM detector for measuring the distortion of modulated carriers. Besides operating instructions, it includes theory of operation, performance checks, adjustments, troubleshooting, parts lists and circuit diagrams.
What this manual covers
- General information: specifications, accessories and options
- Installation
- Operating instructions: distortion measurements, AM carrier distortion (334A only), voltmeter use, meter and output terminals, Option 01 and manual nulling
- Theory of operation
- Maintenance: performance checks, adjustment and calibration procedures, troubleshooting
- Replaceable parts lists and circuit diagrams
Key specifications
- Fundamental frequency range: 5 Hz to 600 kHz, with harmonics measured up to 3 MHz
- Distortion: 0.1% to 100% full scale in 7 ranges
- Automatic nulling; 0.3 V rms input needed for the 100% set level
- Fundamental rejection: more than 80 dB
- Voltmeter: 5 Hz to 3 MHz, 300 µV to 300 V rms full scale
- 334A only: AM detector for carriers from 550 kHz to above 65 MHz
- Option 01: meter with VU response; accessory: HP 10001A 10:1 divider probe
About this document
- Operating and Service Manual, 333A/334A Distortion Analyzers, HP part number 00333-90008, revised October 1975
- 85 pages, English; scanned copy distributed by Agilent, with an errata page added