BK Precision 830 Instruction Manual (PDF)

The B & K-Precision Model 830 Capacitance Meter manual provides detailed instructions for setting up the device, operating it to measure capacitance, and troubleshooting common issues such as zero compensation and overrange conditions. It includes guidance on battery installation, using the autoranging feature, and handling capacitors safely to prevent internal damage to the meter.

  • Brand: BK Precision
  • Model: 830
  • Type: manual
  • File size: 4.0 MB

Manual Overview

The B & K-Precision Model 830 Capacitance Meter manual provides detailed instructions for troubleshooting, calibration, and usage. For troubleshooting, if standard maintenance procedures do not resolve issues, the unit should be securely packed and sent to an authorized service agency with a description of the problem. The meter features a zero adjust control to ensure accurate readings, which should be set to 00.0 without a minus sign. If a minus sign appears, adjust the zero control in the positive direction. The meter automatically ranges to provide the highest resolution, with a selectable range hold feature for faster batch testing of large capacitors. A shorted capacitor will cause an overrange indication, displaying a plus sign and blanking the display. Calibration involves inserting capacitors into the test socket or using test leads for those without leads, ensuring no voltage is applied to the test jacks to avoid damage. The manual also includes a conversion chart for capacitance values and specifies the use of four “C” cells for power, with an option for rechargeable batteries.

Quick Issues Covered

  • Diagnose power-up failures with zero adjust and range settings.
  • Identify error codes via LCD overrange and blank display indicators.
  • Trace circuit schematics for capacitor test socket connections.

Overview

The B & K-Precision Model 830 Capacitance Meter provides a direct reading of capacitance on a 3-3/4 digit LCD display. It features autoranging, automatically selecting the range with the highest resolution, and includes ten full scale ranges from 199.9 picofarads to 199.9 millifarads. The device is suitable for use in electronic engineering labs, production, service shops, and schools. It can check tolerance, sort values, select precision values, measure unmarked capacitors, select matched sets, and measure cable or switch capacitance. The meter is battery-operated, lightweight, and compact, making it portable.

Specifications

  • CAPACITANCE
  • Resolution: 0.1 pF on lowest range
  • Range: 10 ranges, fully automatic with range hold. Full scale value from 199.9 pF to 199.9 mF.
  • Accuracy: AUTO: 0.2% of reading, + 0.5 pF, +1 digit to 19.99 counts. HOLD: All specifications between 180 and 1999 counts same as AUTO. For all readings between 0 and 179 counts add: 0.1% of full scale to 19.99 wF (including 199.9 pF range), 0.5% of full scale from 19.99 uF to 199.9 mF.
  • Reading Update Time: 0.4 – 1.0 second to 20mF, increasing to 6 seconds at 200 mF.
  • Zero Control: Can compensate up-to 25 pF of test lead capacitance, minus (-) sign indicates over-compensation on lowest range.
  • Operating Temperature: +15° to +35° C; usable at reduced accuracy 0° to +50° C.
  • Overrange Indication: Plus (+) sign, with blank display and mF LED on.
  • GENERAL
  • Display: 3-4 digit, 0.5” LCD with low battery indicator.
  • Power Source: 4 standard “C” cells operating from 4.2 to 6 volts, Nicad, alkaline, or zinc-carbon, with provision for ac adapter/charger.
  • Battery Life: 20 hours typical with alkaline batteries.
  • Dimensions: 16x11x6cm (6.4×4.4×2.4”)
  • Weight: 725 g (1.6 lb.)

Controls & Indicators

The front-panel controls/indicators of the B & K-Precision Model 830 Capacitance Meter include a power on-off switch, a range auto-hold switch, a zero adjust control, an LCD display with a low battery indicator, and a capacitor test socket. Additionally, there are test lead jacks and an adapter-charger jack on the unit.

Operation

The B & K-Precision Model 830 Capacitance Meter is used to measure capacitance directly on a 3-3/4 digit LCD display. It features autoranging, which automatically selects the range with the highest resolution, covering ten full scale ranges from 199.9 picofarads to 199.9 millifarads. The range hold feature prevents the range from resetting when a capacitor is removed. The meter is suitable for electronic engineering labs, production, service shops, and schools. It can check tolerance, sort values, select precision values, measure unmarked capacitors, select matched sets, and measure cable or switch capacitance. The device is battery-operated, lightweight, and compact.

To operate the meter, follow these steps:
1. Ensure the meter is equipped with fresh batteries. Install four “C” cells as needed.
2. Set the POWER switch to ON and the RANGE switch to AUTO.
3. Zero the meter by adjusting the ZERO control until the display reads 00.0 without a minus sign.
4. For capacitors with leads, insert them into the test socket. For those without leads, use test leads connected to the “+” and “-” jacks.
5. Read the capacitance value from the LCD display and its unit indicator LED.
6. For faster measurement of similar capacitors, use the RANGE HOLD feature to lock the range.

Note: Fully discharge capacitors before measurement, and observe polarity for polarized capacitors. Never apply voltage to the test jacks to avoid damage.

Download PDF (4.0 MB)