BK Precision 801 Instruction Service Manual Hard to Read Schematics (PDF)

The B&K Model 801 Capacity Analyzer manual provides detailed instructions for setting up, operating, and troubleshooting the device, including tests for capacitor defects, life expectancy, and energy storage capabilities. It also includes specific procedures for performing short, open, leakage, and capacity value tests, both in-circuit and out-of-circuit, to ensure accurate measurements and reliable results.

  • Brand: BK Precision
  • Model: 801
  • Type: service manual
  • File size: 645.0 KB

Manual Overview

The B&K Model 801 Capacity Analyzer service manual provides detailed instructions for troubleshooting, diagnostics, and calibration of the device. The manual emphasizes the importance of factory calibration for 60 cycle AC operation, with specific caution regarding the 100 volt leakage position where 100 volts DC appears on the RED test lead. The troubleshooting flow involves several tests: the Short Test, Open Test, Leakage Test, and Capacity Value Test, each with specific steps for both “in circuit” and “out of circuit” conditions. For instance, the Short Test requires connecting BLUE and BLACK leads to the capacitor and reading the GOOD-BAD scale, while the Open Test involves similar connections but focuses on detecting open circuits. Calibration adjustments are necessary for the Open Test function, which involves adjusting R-122 to set the meter pointer correctly. The manual also provides guidance on testing electrolytic capacitors, highlighting the importance of disconnecting one lead for accurate measurement. Error codes or diagnostics are not explicitly mentioned, but the manual provides clear indications of test results through meter readings.

Quick Issues Covered

  • Diagnoses capacitor shorts and opens in circuit or out of circuit.
  • Calibrates Open's test function using R-122 adjustment.
  • Traces circuits using detailed schematics for troubleshooting.

Overview

The product is a B & K Precision Test Equipment, specifically the Model 801 Capacity Analyzer. The manual covers its functions, which include analyzing capacitors for defects, evaluating life expectancy and energy storing capabilities, measuring capacitor values, testing for shorts and opens, and measuring leakage. The manual provides detailed instructions for operating the device, including specific tests and cautions for use.

Operation

The manual provides instructions for using the B&K Model 801 Capacity Analyzer. It is designed to analyze capacitors for defects affecting electronic equipment, evaluate the life expectancy and energy storing capabilities of electrolytic capacitors, measure the value of other capacitors, test capacitors for shorts and opens, and measure leakage of capacitors. The unit is factory calibrated for 60 cycle AC operation, and caution is advised as 100 volts DC appears on the RED test lead when the selector switch is in the 100 volt leakage position.

For basic usage:

1. Short Test: Rotate the selector to SHORT, connect BLUE & BLACK leads to the capacitor, and read the GOOD-BAD scale.
2. Open Test: Rotate the selector to OPEN, connect BLUE & BLACK leads to the capacitor, and read the GOOD-BAD scale.
3. Leakage Test: Disconnect one capacitor lead if in circuit, rotate the selector to the appropriate LEAKAGE position, connect RED & BLUE leads to the capacitor, adjust the meter for infinity on the leakage scale, depress the test switch, and read leakage resistance on the scale.
4. Capacity Value Test: Disconnect one capacitor lead if in circuit, rotate the selector to the desired capacity range, adjust the calibrate control for full scale, connect BLUE & BLACK leads to the capacitor, and read the value on the scale.
5. Electrolytic Capacitor Test: Disconnect one capacitor lead if in circuit, rotate the selector to CALIBRATE, adjust the calibrate control for full scale, connect BLUE lead to positive and BLACK lead to negative, rotate the selector to the desired range, and read the value on the blue meter scales. If the short lamp glows brightly with no meter reading, the capacitor is shorted. If there is no glow and no reading, the capacitor is open.

Safety & Warnings

CAUTION: When the Selector switch is in the 100 volt leakage position and the Test switch is depressed, 100 volts DC appears on the RED test lead.

Download PDF (645.0 KB)