The BK Precision Model 1823A 2.4 GHz Universal Frequency Counter manual provides detailed instructions on the setup, operation, and troubleshooting of the device, including information on initial inspection, connecting AC power, warming up, and performing various frequency measurements. It also includes safety guidelines, maintenance tips such as fuse replacement and calibration, and the use of RS-232C serial interfacing for enhanced functionality.
Manual Overview
The BK Precision Model 1823A Universal Frequency Counter manual provides detailed guidance on troubleshooting, diagnostics, calibration, and board-level operations. For troubleshooting, the manual emphasizes the importance of initial inspections and ensuring proper connections, particularly for AC power and ventilation. Error codes and diagnostics are facilitated by a self-test mode that checks various operational aspects. Calibration and adjustments are crucial for maintaining accuracy, with specific instructions for fuse replacement and calibration procedures outlined in the maintenance section. The manual provides board-level guidance by detailing the characteristics of inputs A, B, and C, including frequency ranges, sensitivities, and coupling options. It also covers the use of the RS-232C serial interface for data communication. The manual highlights the importance of using the device within specified environmental conditions to ensure precision and reliability. Additionally, safety precautions are emphasized, particularly when working with high voltages, to prevent user hazards and equipment damage.
Quick Issues Covered
- Diagnose power-up failures with pre-heating requirements.
- Adjust and calibrate using detailed procedures in section 4-2.
- Trace circuits using provided schematics for troubleshooting.
Overview
The BK Precision Model 1823A is a 2.4 GHz Universal Frequency Counter designed for high-resolution frequency measurements. It is a microprocessor-controlled instrument that utilizes uniquely developed LSI and expanding/reciprocal systems to achieve high accuracy and sensitivity. The frequency counter covers a range from 0.1 Hz to 2.4 GHz, featuring various measurement functions such as trigger, time interval, frequency ratio, and more. It includes a 9-digit LED display and is suitable for use by scientists, engineers, and technicians due to its precision and reliability. The manual provides detailed instructions on installation, operation, and maintenance of the instrument.
Specifications
- FREQUENCY RANGE: 0.1 Hz to 100 MHz (DC Coupled), 30 Hz to 100 MHz (AC Coupled)
- SENSITIVITY: 0.1 Hz to 1 Hz: 250 mV, 1 Hz to 100 MHz: 30 mV
- COUPLING: AC or DC Selectable
- IMPEDANCE: 1 MΩ Resistance, Shunted by < 40 pF
- ATTENUATOR: x 1 or x 10 Switch Selectable
- LOW PASS FILTER: -3 dB Point of 100 kHz, Switch Selectable
- TRIGGER LEVEL: + 350 mV to – 350 mV (PRESET 0 V)
- SLOPE: Positive or Negative Slope Switch Selectable
- RESOLUTION AND NUMBER OF DISPLAYED DIGIT: Time Base Selector INT | EXT | INT | EXT | INT | EXT | INT | EXT
- Gate Time 0.01S 0.1S 1S 10S
- Number Of Displayed Digit: 5 | 6 6 | 7 7 | 8 8 | 9
- FREQUENCY (Input A,B) RESOLUTION: 0.1 Hz-0.99 Hz 10uHz |1uHz |1uHz |0.1uHz |0.1uHz |10nHz |10nHz |1nHz
- ACCURACY: 0.1Hz to 100 Hz: + Time base error + resolution + 1 count
- PERIOD RANGE: >10nSto10s
- DISPLAY: in. u.m., Sec with decimal point
- TOTAL RANGE: DC to 30 MHz
- CAPACITY: 0 to 999 999 999
- MAX. INPUT VOLTAGE LEVEL: MAX INPUT VOLTAGE (ACV & DCV)
- TIME INTERVAL(A->B): RANGE: 0.1 uSec-10 Sec(0.1 Hz-10 MHz)
- LSD: 100 nSec
- RESOLUTION: + LSD + Trigger Error
- ACCURACY: + LSD + Trigger Error +Time base error x T.
- MULTIPLIER: 1,10,100,1000 (Gate Time: 10 S,1 S,0.1 S,0.01 S)
- RATIO(A/B): RANGE: 0.1 Hz to 10 MHz(input A), 0.1 Hz to 10 MHz(input B)
- RESOLUTION: + LSD +(B Trig. ERROR x FREQ. A)/N
- ACCURACY: +1 COUNT of A+ B TRIG. ERROR x FREQ. A.
- INPUT C. CHARACTERISTICS: FREQUENCY RANGE: 80 MHz to 2.4 GHz
- SENSITIVITY: 25 mV from 80 MHz to 150 MHz, 20 mV from 150 MHz to 2.0 GHz, 60 mV from 2.0 GHz to 2.4 GHz
- COUPLING: AC only
- IMPEDANCE: 50 Ω +5%
- MAX. INPUT LEVEL: 3 Vrms sine wave
- RESOLUTION AND NUMBER OF DISPLAYED DIGIT: Time Base Selector INT | EXT | INT | EXT | INT | EXT | INT | EXT
- Gate Time 0.01S 0.1S 1S 10S
- Number
Safety & Warnings
Please take a moment to read these operating instructions thoroughly and completely before operating this instrument. Pay particular attention to WARNINGS used for conditions and actions that pose hazard to the user and CAUTIONS used for conditions and actions that may damage the instrument.
– Always inspect the instrument and other accessories for any sign of damage or abnormality before every use.
– Never ground yourself and keep your body isolated from ground.
– Never touch exposed wiring, connections, or any live circuit conductors.
– Do not install substitute parts or perform any unauthorized modification to the instrument.
– Use caution when working above 60V DC or 30V AC rms. Such voltages pose a shock hazard.
– Remember that line voltage is present on some power input circuit points such as on-off switches, fuse, power transformers, etc., even when the equipment is turned off.
– Also, remember that high voltage may appear at unexpected points in defective equipment.