The BK Precision manual for models 1710 and 1730 covers setup and operation of the DC power supplies, detailing features such as adjustable voltage and current settings, built-in metering, and safety precautions to prevent electrical shock. It also provides troubleshooting tips, including reverse polarity protection and overload protection to safeguard both the power supply and the equipment being tested.
Manual Overview
The BK Precision Model 1710 and 1730 DC Power Supplies are designed for general-purpose use, providing adjustable outputs of 0-30 volts DC. The Model 1710 offers a current output of 0-1 amp, while the Model 1730 provides 0-3 amps. Both models feature coarse and fine controls for precise voltage and current settings. They include two large panel-mounted meters for continuous monitoring of output voltage and current. These power supplies can operate in either constant voltage or constant current modes, with automatic crossover between the two. LED indicators display the current operational mode. The units are equipped with reverse polarity protection and current limiting to safeguard against improper connections and overloads. The output is isolated from the chassis and earth ground, allowing flexible connection options. Calibration and adjustments are detailed in the manual, ensuring accurate operation. For troubleshooting, the manual provides guidance on error codes and diagnostics, emphasizing safety precautions to prevent electrical shock during testing. The power supplies are suitable for various applications, including service shops, engineering labs, and educational settings.
Quick Issues Covered
- Diagnose power-up failures with LED mode indicators.
- Calibrate Model 1710/1730 for precise voltage/current settings.
- Trace circuits using detailed schematics for troubleshooting.
Overview
The B+K Precision Models 1710 and 1730 are high-quality, general-purpose DC power supplies. Both models provide an adjustable output of 0-30 volts DC, with the Model 1710 offering a current output of 0-1 amp and the Model 1730 offering 0-3 amps. These power supplies feature excellent regulation and low ripple characteristics, with a pre-regulator design to reduce internal power dissipation. They can operate in constant voltage or constant current modes, with smooth and automatic crossover between the two. The units include built-in metering, LED indicators, reverse polarity protection, and overload protection. The output is isolated from chassis and earth ground, allowing flexibility in connections. These power supplies are suitable for various applications, including service shops, engineering labs, production testing, school laboratories, and home use by hobbyists.
Specifications
- OUTPUT VOLTAGE 0 to 30 VDC, coarse and fine adjustment
- OUTPUT CURRENT
- High Range 0 to 3 A (Model 1730), 0 to 1 A (Model 1710)
- Low Range 0 to 0.5 A (Model 1730), 0 to 0.25 A (Model 1710)
- CONSTANT VOLTAGE OPERATION
- Voltage Regulation
- Line (108 – 132 V) 0.01% +3 mV
- Load (no load to full load) 0.01% +3 mV
Operation
The manual provides detailed operating instructions for the B+K Precision Model 1710 and 1730 DC Power Supplies. These power supplies are designed for general-purpose use, offering 0-30 volts DC output with adjustable current settings. The Model 1710 provides a current output of 0-1 amp, while the Model 1730 offers 0-3 amps. They feature both constant voltage and constant current operation modes, with smooth and automatic crossover between the two. Users can preset current limits, and the units will regulate either voltage or current based on load variations. The power supplies include reverse polarity protection and current limiting to safeguard both the equipment and the power supply. The output is isolated from chassis and earth ground, allowing flexible connections. Additionally, two supplies can be connected in series or parallel for increased voltage or current output. The manual also highlights safety precautions, such as using a polarized 3-wire AC outlet and avoiding unnecessary exposure to high voltage.
Safety & Warnings
Normal use of test equipment can expose users to electrical shock hazards, especially when testing in environments with exposed high voltage. A current of 10 milliamps passing through the heart can be fatal, and voltages as low as 35 volts can be dangerous. Users should adopt safe work habits to avoid contact with high voltage and steer current away from the heart. Specific precautions include:
1. The DC output of the power supply poses little shock risk, but certain conditions can create hazards:
a. A “hot chassis” type equipment under test can be hazardous unless unplugged or isolated.
b. If the equipment under test is powered and uses high voltage, the power supply outputs may float to high potential.
c. Discharge high-voltage capacitors in equipment before testing if they use high voltage during normal operation.
2. Use a polarized 3-wire AC outlet to ensure the power supply is properly grounded.
3. Avoid unnecessary exposure to high voltage. Only remove housings when necessary, turn off equipment when making high-voltage connections, and discharge capacitors after power removal.