BK Precision 3010 Instruction Maintenance Manual (PDF)

The B&K-PRECISION Model 3010 Function Generator manual provides detailed instructions on setting up, operating, and troubleshooting the device, including frequency and waveform selection, DC output, and voltage-controlled operation. It also covers safety precautions to prevent electrical shock during testing, maintenance, calibration procedures, and warranty service instructions.

  • Brand: BK Precision
  • Model: 3010
  • Type: manual
  • File size: 7.6 MB

Manual Overview

The B&K Precision Model 3010 Function Generator manual provides detailed guidance on the safe operation, maintenance, and calibration of the device. It emphasizes the importance of safety precautions when using test equipment due to the risk of electrical shock, advising users to avoid unnecessary exposure to high voltage and to employ isolation transformers with transformerless equipment. The manual outlines the function generator’s specifications, including its frequency range of 0.1 Hz to 1 MHz and various output waveforms such as sine, square, triangle, and TTL. It provides instructions on frequency and waveform selection, as well as adjustments for DC offset and amplitude control. The calibration section details the procedure for maintaining accuracy, with emphasis on dial accuracy and output impedance. The manual also includes a troubleshooting flow, focusing on error codes and diagnostics to assist technicians in identifying and resolving issues at the board level, ensuring the function generator operates within specified parameters.

Quick Issues Covered

  • Diagnose Model 3010 power-up failures and calibration issues.
  • Identify and resolve error codes specific to Model 3010.
  • Use schematics for circuit tracing in Model 3010 diagnostics.

Overview

The B & K-PRECISION Model 3010 Function Generator is a versatile signal source suitable for engineering, industrial, educational, and hobbyist applications. It offers a wide frequency range from 0.1 Hz to 1 MHz, covering subaudible, audio, ultrasonic, and RF applications. The device features sine, square, triangle, and TTL outputs, with a continuously variable DC offset control and external VCO control for enhanced versatility. The manual includes sections on specifications, operating instructions, applications, theory of operation, maintenance and calibration, and warranty service instructions.

Specifications

  • Basic Outputs: Square, sine, triangle, DC and TTL Square Wave
  • Frequency Range: 0.1 Hz to 1 MHz in six ranges
  • External Frequency Control: VCO; greater than 100:1 on any frequency range (linear)
  • Maximum Input: +20 V peak
  • Input Impedance: 1000 ohms, nominal
  • Dial Accuracy: 5% of full scale to 500 kHz, 8% of full scale from 500 kHz to 1 MHz
  • Dial Range: Greater than 100:1
  • Output Impedance: 600 ohms, ±5%
  • Output Amplitude: 20 V p-p open circuit; 10 V p-p into 600 ohms
  • Amplitude Control: Continuously variable, greater than 30 dB range
  • DC Offset: Continuously variable, ±10 V, open circuit; ±5 V, into 600 ohms
  • Sine Wave Distortion: Less than 1%, 1 Hz to 100 kHz
  • Square Wave Non-Symmetry: Less than 1%, to 100 kHz
  • Triangle Wave Non-Linearity: Less than 1%, to 100 kHz
  • Square Wave Rise/Fall Time: Less than 100 nSEC at maximum output amplitude
  • Sine Wave Amplitude Flatness: ±.3 dB to 1 MHz at maximum output amplitude
  • TTL Square Wave Response: Less than 25 nSEC rise/fall time
  • Short-Term Stability: 0.5%
  • Power Requirements: 105 to 130 VAC, 60 Hz. 8 watts max.
  • Dimensions (HWD): 3.2×11.3×7.7"
  • Weight: 2 lbs., 9 oz. without line cord; 2 lbs., 14 oz. with line cord
  • Handle: Four positions: integral part of case

Controls & Indicators

– POWER on-off switch: Depressing this button turns the 3010 on. To turn off, push again and release.
– RANGE selectors: Decade frequency type. Multiplying the range selected times the FREQUENCY dial indication gives the output frequency, which applies for all functions.
– FUNCTION selectors: Select square, sine, or triangle output waveform which appears at 600 ohms OUTPUT jack.
– AMPLITUDE control: Controls the amplitude of the output signal, which appears at 600 ohms jack. This control does not apply to the DC OFFSET voltage or to the TTL output.
– DC OFFSET control: Adds positive or negative DC component to the signal appearing at 600 ohms OUTPUT jack. Continuously variable for +5 V (600 ohms) or ±10 V (open circuit).

Safety & Warnings

Normal use of test equipment can expose users to electrical shock hazards, particularly when testing in areas with exposed voltage. Even voltages as low as 35 volts DC or AC RMS can be dangerous and potentially lethal. To reduce risk, the following safety precautions should be observed:

1. Avoid unnecessary exposure to high voltage. Only remove housings and covers when necessary, and ensure equipment is turned off when making test connections in high-voltage circuits. Discharge high-voltage capacitors after removing power.
2. Use insulated floor material or a large, insulated floor mat to stand on, and an insulated work surface for equipment. Ensure these surfaces are not damp or wet. If insulated flooring is unavailable, wear heavy gloves.
3. Employ the “one hand in the pocket” technique when handling an instrument probe to avoid contact with nearby metal objects that could provide a ground return path.
4. Always use an isolation transformer to power transformerless “hot chassis” equipment, such as recent television sets and audio equipment, to prevent dangerous shock hazards and potential damage to test instruments.
5. Use only a 3-wire outlet for test instruments or equipment with a 3-wire AC power plug to maintain safety by keeping exposed elements at earth ground.
6. Familiarize yourself with the equipment being tested and the location of its high voltage points, but remember that high voltage may appear unexpectedly in defective equipment.
7. Be aware that AC line voltage may be present on some power input points, such as on-off switches, fuses, and power transformers, even when equipment is turned off.
8. Never work alone. Ensure someone is nearby to provide aid if necessary. CPR training is highly recommended.