BK Precision 2625 Instruction Manual (PDF)

The manual for B+K Precision Spectrum Analyzers models 2615A, 2620A, 2625, and 2630 covers the setup, operation, and troubleshooting of these devices, including safety precautions to prevent electrical shock and damage to equipment. It provides detailed specifications, calibration instructions, and guidance on frequency measurements, sensitivity, and frequency response, as well as information on optional accessories and warranty service instructions.

  • Brand: BK Precision
  • Model: 2625
  • Type: manual
  • File size: 1.7 MB

Manual Overview

The manual for BK Precision Spectrum Analyzers, models 2615A, 2620A, 2625, and 2630, provides detailed guidance on troubleshooting, error diagnostics, calibration, and board-level adjustments. Troubleshooting involves understanding the specifications and operational features of each model, such as frequency range, input impedance, and resolution bandwidth. Error diagnostics are facilitated through the identification of error codes and the use of diagnostic tools to assess performance issues. Calibration is crucial for maintaining accuracy, with vertical and horizontal calibration procedures outlined to ensure precise measurements. Board-level guidance includes handling and adjustments of input attenuators and frequency stability checks. The manual emphasizes safety precautions, particularly when dealing with high voltage and “hot chassis” equipment, recommending the use of isolation transformers to mitigate shock hazards. The document also highlights the importance of using insulated materials and maintaining proper grounding to prevent electrical accidents.

Quick Issues Covered

  • Diagnose power-up failures with step-by-step troubleshooting guide.
  • Identify error codes specific to Models 2615A, 2620A, 2625, 2630.
  • Calibration procedures for vertical and horizontal adjustments.

Overview

The document provides information about the B&K Precision Models 2615A, 2620A, 2625, and 2630 Spectrum Analyzers. It includes sections on test instrument safety, specifications, optional accessories, general information, introduction to spectrum analysis, controls and indicators, calibration, and customer support. The manual covers safety precautions, operating conditions, and maintenance instructions. Specifications detail frequency input, tracking generator, amplitude, and general characteristics such as display and power consumption.

Specifications

  • Frequency Input
  • Frequency range: 0.15 MHz to 1050 MHz (Models 2625 and 2630)
  • Frequency range: 0.15 MHz to 500 MHz (Models 2615A and 2620A)
  • Input impedance: 50Ω
  • Input connector: BNC
  • Input attenuator: 0 to 40 dB (4 x 10 dB steps)
  • Input attenuator accuracy: ±1 dB/10 dB step
  • Center frequency display accuracy: ±100 kHz
  • Max. input level: +10 dBm, 25VDC (0 dB attenuation)
  • Marker accuracy: ±(0.1% span + 100 kHz)
  • Frequency display resolution: 100 kHz
  • Frequency scanwidth: 100 kHz/div to 100 MHz/div (Models 2625 and 2630)
  • Frequency scanwidth: 50 kHz/div to 50 MHz/div (Models 2615A and 2620A)
  • Frequency scanwidth accuracy: ±10%
  • Frequency stability: Drift <150 kHz/hour
  • IF-Bandwidth (-3 dB): Resolution 400 kHz and 20 kHz (Models 2625 and 2630)
  • IF-Bandwidth (-3 dB): Resolution 250 kHz and 20 kHz (Models 2615A and 2620A)
  • Video-Filter on: 4 kHz
  • Amplitude
  • Amplitude range: -100 dBm to +13 dBm
  • Screen display range: 80 dB (10 dB/div.)
  • Reference level: -27 dBm to +13 dBm (in 10 dB steps)
  • Reference level accuracy: ±2 dB
  • Average noise level: -99 dBm (20 kHz BW) (Models 2625 and 2630)
  • Average noise level: -99 dBm (20 kHz BW) (Models 2615A and 2620A)
  • Distortion: <-75 dBc: 2nd and 3rd harmonic
  • 3rd order intermodulation: -70 dBc (two signals >3 MHz apart)
  • Sensitivity: dB above average noise level
  • Log Scale fidelity: ±2 dB (without atm.) Ref.: 250 MHz
  • IF gain: 10 dB adjustment range
  • General
  • Display: CRT, 6 inch, 8 x 10 div. internal graticule
  • Trace rotation: Adjustable on front panel
  • Output Probe Power: 6V
  • Line voltage: 115/230V ±10%, 50-60Hz
  • Power consumption: approx 27W
  • Operating ambient temperature: +10°C to +40°C
  • Protective system: Safety Class I (IEC 1010-1)
  • Weight: Approx. 15.4 lb (6 kg) (Models 2625 and 2630)
  • Weight: Approx. 13.2 lb (5 kg) (Models 2615A and 2620A)
  • Dimensions: 4.9 in. (125 mm) H x 11.2 in. (285 mm) W x 15 in. (380 mm) D
  • Accessories Supplied
  • Power Cord
  • Instruction Manual

Safety & Warnings

Normal use of test equipment can expose users to electrical shock, especially when testing where exposed voltage is present. A current of 10 milliamps passing through the heart can be lethal, and voltages as low as 35 volts can be dangerous. To reduce risk, adhere to safety precautions:

1. Avoid unnecessary exposure to high voltage. Remove housings and covers only when necessary and turn off equipment while making test connections in high-voltage circuits. Discharge high-voltage capacitors after removing power.
2. Familiarize yourself with the equipment and its high voltage points, but be aware that high voltage may appear unexpectedly in defective equipment.
3. Use insulated floor material or a large insulated mat to stand on, and ensure work surfaces are not damp or wet.
4. Use the “one hand in the pocket” technique when handling an instrument probe to avoid providing a ground return path.
5. Remember that ac line voltage is often present on power input circuits even if the equipment is turned off.
6. For equipment with a two-wire ac power cord, use an isolation transformer to prevent shock hazards and damage to test instruments.
7. Use only a 3-wire outlet for test instruments or equipment with a 3-wire power plug to keep exposed elements at earth ground.
8. B+K Precision products are not authorized for use involving direct contact with the human body or as critical components in life support devices.
9. Never work alone; someone should be nearby to render aid if necessary. Training in CPR and first aid is recommended.

Download PDF (1.7 MB)