Tektronix-p6022-part1-operation-service MANUAL – Service manual (PDF)

The Tektronix P6022 Current Probe manual provides detailed instructions on the setup, operation, and troubleshooting of the probe, including its use with an oscilloscope and termination to measure alternating current waveforms. It covers topics such as increasing sensitivity by adjusting the number of turns through the probe, handling high currents safely, and minimizing interference from external magnetic fields.

  • Brand: Tektronix-p6022-part1-operation-service
  • Model: MANUAL
  • Type: service manual
  • File size: 23.4 MB

Overview

The Tektronix P6022 Current Probe is designed to measure alternating current waveforms and is used in conjunction with devices like the Type 134 Current Probe Amplifier, 7A14 Current Probe Amplifier plug-in, or the P6022 Passive Termination to view and measure current waveforms on an oscilloscope. The manual primarily covers the use of the P6022 Current Probe with the Termination, while additional information for use with current probe amplifiers can be found in the respective amplifier instruction manuals. The manual includes specifications, electrical characteristics, environmental and physical characteristics, and operational instructions for the P6022 Current Probe.

Specifications

  • Sensitivity One milliampere or ten milliamperes for each millivolt at oscilloscope input, selected by slide switch.
  • Accuracy Within +3%
  • Bandwidth (with Type 454 test oscilloscope) 1 mA/mV 8.5 kilohertz or less to 100 megahertz or more
  • Bandwidth (with Type 454 test oscilloscope) 10 mA/mV 935 hertz or less to 120 megahertz or more
  • Maximum CW Current 6 amperes peak to peak sinewave between 10 kHz and 10 MHz at 1mA/mV; between 3kHz and 10 MHz at 10mA/mV
  • Maximum Pulse Current 100 amperes peak, not to exceed 9 ampere-microseconds or 2 amperes RMS
  • Maximum Voltage 600 volts (DC + peak AC)
  • Maximum DC Saturation 200 milliamperes
  • Delay Time 9 nanoseconds (5 foot probe) or 15 nanoseconds (9 foot probe) with termination.
  • Insertion Impedance 0.03 ohms or less at 1 MHz, increasing to 0.2 ohms or less at 120 MHz.
  • Temperature Storage —40°C to +65°C
  • Temperature Operating 0°C to +50°C
  • Altitude Storage To 50,000 feet
  • Altitude Operating To 15,000 feet
  • Vibration Operating 15 minutes along each axis at 0.015". Vary the frequency from 10 to 50 to 10c/s in 1-minute cycles.
  • Shock Nonoperating 30 g’s, 1/2 sine, 11 microseconds duration, 2 shocks per axis. Total of 6 shocks.
  • Dimensions Probe Body 0.6" H X 0.4" WX 6.0" L
  • Dimensions Probe Cable 57.5" L (5 foot cable), or 105.5” L (9 foot cable)
  • Dimensions Termination 0.9"HX1.07 WX 3.5"L
  • Weight P6022 Probe and Cable 2.5 oz.
  • Weight Termination 2.0 oz.
  • Finish Termination Plated metal castings with gray plastic covers.

Controls & Indicators

Refer to the PDF for details.

Installation / Setup

When installing the P6022 Current Probe, it is necessary to use an oscilloscope with a vertical input impedance of one megohm. To increase the current sensitivity, the number of turns passing through the core of the probe can be increased, effectively forming a multi-turn primary winding. Care should be taken not to leave the probe clamped around a conductor when disconnecting the probe cable from the termination, as this can cause high voltage in the secondary winding and potentially damage the probe transformer. The probe is shielded to minimize interference from external magnetic fields, but strong fields may still affect measurements. In such cases, using two probes with a differential-input oscilloscope can help minimize interference.

Operation

The P6022 Current Probe, when used with a termination and an oscilloscope, is designed for measuring alternating current waveforms. An oscilloscope with a vertical input impedance of one megohm is required for its operation. To increase the current sensitivity, the number of turns passing through the core of the probe can be increased, which increases the secondary current. The probe is shielded to minimize interference from external magnetic fields, but strong fields may still affect measurements. In such cases, using two current probes and a differential-input oscilloscope can help minimize interference. When measuring high currents, the probe should not be left clamped around the conductor while disconnecting the probe cable from the termination to avoid damage. The insertion impedance of the probe should be considered when observing fast-rise signals.

Troubleshooting & Error Codes

Refer to the PDF for details.

Safety & Warnings

The manual does not contain any explicit safety warnings or precautions.

Download MANUAL Service manual (PDF – 23.4 MB)

Tektronix P6022 Manual: Provides instructions for operation, troubleshooting, and calibration. Essential for service and maintenance tasks.

  • Brand: Tektronix-p6022-part1-operation-service
  • Model: MANUAL
  • Type: Service manual
  • File size: 23.4 MB

Overview

Brief overview of the product.

Specifications

  • Key spec 1
  • Key spec 2

Controls & Indicators

Control panel/button layout.

Installation / Setup

  1. Step 1…
  2. Step 2…

Operation

Basic usage, modes, cycles.

Troubleshooting & Error Codes

Issue / Code Fix
Example How to fix…