The manual for the Hickok 539B tube tester outlines the procedure for testing germanium diodes, focusing on forward and reverse conduction tests by connecting the diode to specific pins on the octal test socket. It provides specific settings and acceptable reading limits for various diode models to determine if they are functioning correctly.
Manual Overview
The Hickok 539B tube tester manual outlines procedures for testing germanium diodes through two primary tests: forward conduction and rectification, followed by reverse conduction. For the forward conduction and rectification test, the diode’s cathode should be connected to octal test socket pin No. 3, and the other end to pin No. 6. If the meter reads backwards, the diode should be reversed. This setup also applies to the reverse conduction test. Specific settings for each diode type are provided, including bias, tube selectors, and shunt values. Acceptable readings on the 6000 scale vary by diode type: IN34 should read less than 800, IN38 less than 350, IN48 less than 1400, IN91 less than 2000, and IN93 less than 300 for the reverse conduction test. These readings indicate the diode is functioning correctly. The manual does not provide specific troubleshooting flow, error codes, or board-level guidance beyond these testing procedures.
Quick Issues Covered
- Diagnose diode forward and reverse conduction issues.
- Calibrate using specific voltage and shunt settings for each diode type.
- Verify diode health with meter readings on 6000 scale.
Overview
Testing germanium diodes on the Model 539B tube tester involves two tests: forward conduction and rectification, and reverse conduction. The procedure requires connecting the diode to specific pins on the octal test socket and observing meter readings to determine if the diode is functioning correctly.
Specifications
Refer to the PDF for details.
Controls & Indicators
Refer to the PDF for details.
Installation / Setup
Refer to the PDF for details.
Operation
The operation involves testing germanium diodes using the Model 539B Tube Tester through two tests: Forward Conduction and Rectification, and Reverse Conduction. For the Forward Conduction and Rectification test, connect the cathode end of the diode to octal test socket pin No. 3 and the other end to pin No. 6. If the meter reads backwards, reverse the diode. This setup is also used for the Reverse Conduction test. The diode is considered OK if the reading is below a specified value on the 6000 scale, which varies depending on the diode type.
Troubleshooting & Error Codes
Refer to the PDF for details.
Safety & Warnings
Refer to the PDF for details.
Download DATA Manual (PDF – 45.7 KB)
Hickok 539B Germanium Diode Test Manual: Provides instructions for testing and troubleshooting germanium diodes. File size: 45.7 KB.
- Brand: Hickok-539b-germanium-diode-test
- Model: DATA
- Type: Manual
- File size: 45.7 KB
Overview
Brief overview of the product.
Specifications
- Key spec 1
- Key spec 2
Controls & Indicators
Control panel/button layout.
Installation / Setup
- Step 1…
- Step 2…
Operation
Basic usage, modes, cycles.
Troubleshooting & Error Codes
| Issue / Code | Fix |
|---|---|
| Example | How to fix… |
Safety & Warnings
- Important warning…
Downloads
More manuals from Hickok
- Hickok-tv-10a-10u-operation-service-manual-navships 93069 – Service manual (PDF)
- Hickok-tv-10-10u-operation-service-manual-navships 93021 – Service manual (PDF)
- Hickok-tv-7-operation-manual-supplement-to33aa21-5-31m 1973 – Operation manual (PDF)
- Hickok-tv-7-operation-manual-supplement-tb11-5083-1 1953 – Operation manual (PDF)
- Hickok-tv-7-operation-service-manual-tm11-5083 1953 – Service manual (PDF)