BK Precision 4003 Instruction Manual (PDF)

The manual for BK Precision Models 4001 and 4003 covers setup instructions, operation guidelines, and troubleshooting tips for the 3 MHz function generator, including details on unpacking, environmental conditions, and voltage settings. It also provides specific operational instructions for each model, safety precautions, and maintenance advice such as cleaning and fuse replacement.

  • Brand: BK Precision
  • Model: 4003
  • Type: manual
  • File size: 1.4 MB

Manual Overview

The BK Precision Models 4001 and 4003 Function Generators are designed for a range of applications, offering six waveform outputs including sine, square, triangle, ramp, and pulse forms. The manual provides detailed instructions for operation, emphasizing the importance of proper environmental conditions and voltage settings to prevent damage. Troubleshooting guidance includes checking fuse integrity if the device fails to power on, with specific fuse ratings provided for different voltage settings. Calibration and adjustments are crucial for maintaining waveform accuracy, with specifications for frequency stability and waveform symmetry detailed for each output type. The manual advises regular maintenance, including cleaning and recalibration, to ensure optimal performance. It also highlights safety precautions, such as avoiding external voltage application to output connectors and ensuring proper grounding. Board-level guidance is not explicitly detailed, but the manual stresses that servicing should be conducted by qualified technicians to avoid compromising the instrument’s safety and functionality.

Quick Issues Covered

  • Diagnose power-up failures by checking fuse and voltage settings.
  • Calibrate waveform symmetry using front panel controls.
  • Use schematics for circuit tracing and troubleshooting.

Overview

The BK Precision Models 4001 and 4003 are 3 MHz function generators. The manual provides detailed instructions and specifications for both models, covering aspects such as general description, specifications, front panel description, and operation instructions. The manual also includes safety precautions, maintenance guidelines, and fuse replacement instructions.

Specifications

  • Frequency : 0.5Hz ~3Mhz in 6 steps, controlled by rotary switch.
  • Output wave form : Sine, Square, Triangle, Ramp, Positive Pulse and Negative Pulse, 6 wave forms.
  • Stability : 0.1% – 15 minutes after switch on.
  • 0.2% – 24 hours after switch on.
  • DC offset : Continuously Adjustable, +10V at non-load or ±5V at 50Ω load.
  • Limits of operation : 0°C ~40°C, 10 ~80% R.H.
  • Storage environment : -20°C ~70°C, 0 ~90% R.H.
  • Power source : AC 115V (±10%) 50/60Hz, FUSE : 600mA
  • AC 230V (±10%) 50/60Hz, FUSE : 300mA
  • Power consumption : 25W
  • Dimensions : 27.5 x 9.0 x 30.0cm
  • Weight : 2.5Kgs Net.
  • Accessory : Power cord, manual.
  • RAMP WAVE Frequency : 0.5Hz -2.5MHz
  • Symmetry : 80% (rise wave) to 20% (fall wave) < 5%, 1Hz ~100KHz.
  • Rise Wave Linearity : <2%, 1Hz- 100KHz.
  • TRIANGLE WAVE Frequency : 0.5Hz-3MHz.
  • Symmetry : 50% (rise wave) to 50% (fall wave), <2%, 1Hz ~100KHz.
  • Linearity : <1%, 1Hz ~100KHz.
  • SINE WAVE Frequency : 0.5Hz ~3MHz.
  • Distortion : <2%, 1Hz ~100KHz.
  • Harmonic ratio : <30dB, 100KHz ~3MHz.
  • Frequency Response : <0.1dB up to 100KHz.
  • <1dB 100KHz to 3MHz.
  • SQUARE WAVE Frequency : 0.5Hz ~3MHz.
  • Symmetry : 50% (positive half) to 50% (negative half), <2%, 1Hz ~100KHz.
  • Rise time : <60ns.

Installation / Setup

The installation/setup highlights for the BK Precision Models 4001 and 4003 Function Generators include the following:

1. Unpacking and Inspection: Upon receipt, immediately unpack and inspect the instrument for any damage or missing accessories. Report any issues to the dealer promptly.

2. Environmental Conditions: The operational temperature range is 10°C to 40°C (50°F to 104°F). Avoid using the instrument in strong magnetic or electric fields as they may interfere with measurements.

3. Line Voltage Check: Ensure the line voltage selector plug is set to the correct position for the available AC line voltage before connecting the power plug. The instrument can operate on 115V or 230V, and the corresponding fuses should be used (600mA for 115V and 300mA for 230V).

4. Safety Precautions:
– Use only a 3-wire outlet for grounding.
– Do not obstruct ventilation holes.
– Never apply external voltage to the output BNC or excessive voltage to the input BNC.
– The instrument is designed for indoor use only and falls under Installation Category II, Pollution Degree 2.

5. General Handling: Avoid placing heavy objects on the instrument, and do not use it as a surface for hot items like soldering irons. Avoid inserting metal objects into the ventilation fan and refrain from moving the instrument by pulling on the power or probe cords.

6. Calibration and Maintenance: Regular cleaning and recalibration are recommended to maintain the instrument’s appearance and functionality. Use a soft cloth with a mild cleaning solution for cleaning. If inoperative, refer servicing to a qualified repair facility.

7. Fuse Replacement: If the fuse blows, replace it with the correct value fuse based on the line voltage. The fuseholder is located on the rear panel, and proper installation is necessary to ensure correct line voltage selection.

Safety & Warnings

Unpack the instrument immediately upon receipt and inspect for any damage or shortage of accessories. Notify the dealer if any issues are found. Operate the instrument within the temperature range of 10°C to 40°C to avoid circuit damage. Avoid using the instrument in areas with strong magnetic or electric fields. Ensure the line voltage selector plug is set correctly before connecting to an AC line outlet. Incorrect voltage can damage the instrument. Follow these operational hints: do not place heavy objects on the instrument, avoid placing hot soldering irons near it, do not insert metal objects into the ventilation fan, and do not move the instrument by pulling the power or probe cords. Use the instrument as specified by the manufacturer to ensure protection is not impaired. To prevent electric shock, only qualified personnel should use the instrument for testing, do not operate with the cover removed unless qualified, use a 3-wire outlet for grounding, do not block ventilation holes, never apply external voltage to the output BNC, avoid excessive voltage to the input BNC, and use the instrument indoors only. The equipment is evaluated for Installation Category II, Pollution Degree 2.

Download PDF (1.4 MB)