BK Precision 1856D Instruction Manual (PDF)

The BK Precision Model 1856D 3.5GHz Frequency Counter manual provides detailed instructions for setup, operation, and troubleshooting, including initial inspection, connecting AC power, and using controls and indicators for frequency, period, total, and RPM measurements. It also covers maintenance tasks such as fuse replacement, adjustment, calibration, and cleaning, while emphasizing safety precautions and the importance of using the equipment within specified temperature and humidity ranges.

  • Brand: BK Precision
  • Model: 1856D
  • Type: manual
  • File size: 196.6 KB

Manual Overview

The BK Precision Model 1856D 3.5GHz Frequency Counter manual provides detailed guidance for troubleshooting, error diagnostics, calibration, and board-level adjustments. The frequency counter is a microprocessor-controlled instrument designed for high accuracy frequency measurements with a range from 0.1 Hz to 3.5 GHz. It features a 9-digit LED display and supports RPM, period, and total measurements. For troubleshooting, the manual emphasizes the importance of using the correct input settings and checking the self-test function to diagnose operational issues. Error codes and diagnostics are not explicitly detailed, but users are advised to ensure proper connection and configuration of inputs and outputs. Calibration and adjustments involve using the internal or external time base selector to maintain accuracy, with specific gate times affecting resolution. Board-level guidance includes ensuring proper impedance matching and using the appropriate attenuator settings. The manual also highlights the importance of maintaining the device within specified environmental conditions to ensure optimal performance.

Quick Issues Covered

  • Diagnose power-up failures with warm-up period and power cord checks.
  • Calibrate frequency measurements using internal/external time base settings.
  • Trace circuits with detailed schematics for Input A and C characteristics.

Overview

The BK Precision Model 1856D is a 3.5GHz frequency counter designed for frequency measurement. It is a microprocessor-controlled instrument featuring high accuracy with a 9-digit resolution and a frequency range from 0.1 Hz to 3.5 GHz. The device includes features such as RPM measuring function, external frequency standard input, attenuator, period, total, low pass filter, and line filter. It is equipped with a self-test function for quick checks of operating modes, which can be selected via a front panel push button switch. The product is lightweight and compact, making it suitable for use by scientists, engineers, experimenters, communications technicians, hobbyists, or field technicians.

Specifications

  • FREQUENCY RANGE: 0.1 Hz to 100 MHz (DC coupled), 30 Hz to 100 MHz (AC coupled)
  • SENSITIVITY: 30 mV rms
  • COUPLING: AC or DC selectable
  • IMPEDANCE: 1 MQ resistance shunted by < 40 pF
  • ATTENUATOR: X1 or x10 switch selectable
  • LOW PASS FILTER: -3 dB point at approx. 100 KHz, switch selectable
  • ACCURACY: 0.1Hz to 999Hz + Time base stability + 1 count + 1Hz; 1KHz to 99KHz + Time base stability + 1 count + 0.1Hz; 100KHz to 100MHz + Time base stability + res. (Table 1 x 10)
  • PERIOD RANGE: 10 nS to 10$
  • DISPLAY: n/u/m (nano/micro/milli) Sec. with dec point
  • TOTAL RANGE: 10 Hz to 30 MHz
  • CAPACITY: 0 to 999 999 999, OVERFLOW: “OF”
  • RPM RANGE: 600 to 600 x 10° RPM, OVERFLOW: “OF”
  • FREQUENCY RANGE (Input C): 80 MHz to 3.5 GHz
  • SENSITIVITY (Input C): 15mV from 80 MHz to 2000 MHz; 20mV from 2000 MHz to 3.0 GHz; 30mV from 3.0GHZ to 3.2GHz; 50mV from 3.2GHz to 3.5GHz
  • COUPLING (Input C): AC only
  • IMPEDANCE (Input C): 50 QO +5%
  • MAX. INPUT LEVEL (Input C): 3 Vrms sine wave
  • TIME BASE TYPE: TCO (Temperature Controlled Oscillator)
  • TIME BASE FREQUENCY: 10.000000 MHz
  • TIME BASE STABILITY: +1 PPM (+1 count)
  • LINE VOLTAGE STABILITY: Less than + 1 PPM with + 10% line voltage variation
  • TEMPERATURE STABILITY: +5 PPM from 0°C to 50°C
  • MAX. AGING RATE: +5 PPM/year
  • INT. STD. OUT: 10 MHz (Internal Standard Frequency Output)
  • LEVEL: 1 V p-p or more. TTL (10TTL)
  • CAPACITANCE: 15pF
  • EXT. STD. IN: 10 MHz (External Standard Frequency Input)
  • LEVEL: 1.5Vrms to 5Vrms
  • IMPEDANCE: Approx. 600Q
  • DISPLAY: Nine Digit 0.56" LED with M (Mega) or K (Kilo) Hz; nano, micro, or milli (n/u/m) Sec.; GT (GATE TIME); H (HOLD); and OF (OVERFLOW) indicators

Safety & Warnings

Before every use, inspect the instrument and other accessories for signs of damage or abnormality. Keep your body isolated from ground. Never touch exposed wiring, connections, or any live circuit conductors. Do not install substitute parts or perform any unauthorized modifications on the instrument. Use caution when working above 60V DC or 30V AC rms, as such voltages pose a shock hazard. Remember that line voltage is present on circuit points such as on-off switches, fuse, power transformers, etc., even when the equipment is turned off. High voltage may appear at unexpected points in defective equipment.

Download PDF (196.6 KB)