BK Precision 1405 Instruction Manual (PDF)

The manual for the B&K Model 1405 Oscilloscope provides detailed instructions for its setup, operation, and troubleshooting, including safety precautions to prevent electrical shock during use. It covers features such as vertical and horizontal amplifier specifications, sweep characteristics, and power requirements, along with circuit descriptions and maintenance guidelines.

  • Brand: BK Precision
  • Model: 1405
  • Type: manual
  • File size: 1.8 MB

Manual Overview

The B&K Model 1405 Oscilloscope manual provides detailed guidance on the operation, troubleshooting, and maintenance of this device. The oscilloscope features a 3-inch cathode ray tube with a vertical-axis sensitivity better than 10 mV/DIV and a frequency response from DC to 5 MHz. It employs DC amplifiers for both vertical and horizontal axes, utilizing all-transistorized circuitry for low power consumption and heat generation. The manual emphasizes safety precautions due to the inherent risks of electrical shock during testing. Troubleshooting involves understanding the circuit description, which includes a vertical circuit with a differential amplifier and a horizontal circuit with a saw-tooth generator and amplifier. Calibration and adjustments are facilitated through attenuators and gain controls. The power supply circuit provides necessary voltages for operation. For diagnostics, the manual suggests familiarizing oneself with the equipment’s high-voltage points and using isolation transformers to prevent shock hazards. The manual also includes a block diagram and external view for reference, aiding in board-level guidance and ensuring proper handling and operation of the oscilloscope.

Quick Issues Covered

  • Diagnose vertical amplifier issues with sensitivity below 10 mV/DIV.
  • Error codes for power supply circuit failures, including zener diode issues.
  • Schematics for circuit tracing of direct-coupled differential amplifiers.

Overview

The B & K Model 1405 Oscilloscope is a highly sensitive and stable device featuring a 3-inch cathode ray tube. It is designed for easy operation and includes features such as vertical-axis sensitivity better than 10 mV/DIV, a frequency response from DC to 5 MHz, and all transistorized circuitry for low power consumption. The oscilloscope is compact, lightweight, and easily portable. The manual covers features, specifications, circuit description, operating instructions, applications, handling cautions, adjustments, and maintenance.

Specifications

  • CATHODE RAY TUBE: 75AVB31
  • VERTICAL AMPLIFIER:
  • Deflection sensitivity: 10 mV/DIV or better
  • Frequency response: DC to 5 MHz (-3 dB)
  • Input impedance: 1 MΩ shunted by 35 pF max
  • Overshoot: 5% or less
  • Attenuator: 1, 1/10, 1/100 multiplier within ±5%
  • Gain control range: Continuously variable range greater than 22 dB
  • Rated maximum input voltage: 300 VDC + AC peak or 600 Vp-p
  • HORIZONTAL AMPLIFIER:
  • Deflection sensitivity: 300 mV/DIV or better
  • Frequency response: DC to 250 kHz with EXT. GAIN Control set at maximum
  • Input impedance: 1 MΩ (+20%) shunted by 30 pF max. (SYNC-INT)
  • Attenuator (EXT. GAIN): Continuously variable to zero
  • Rated maximum input voltage: 100 Vp-p
  • SWEEP CHARACTERISTICS:
  • Sweep frequency: 10 Hz to 100 kHz continuously variable in 4 ranges
  • Sweep linearity: Within 5%
  • Synchronizing: Internal or external negative
  • Signal amplitude requirement for synchronization: Internal: more than 1 DIV deflection on cathode ray tube screen, External: more than 2 Vp-p
  • INTENSITY MODULATION:
  • Required signal: 25 Vp-p
  • DIRECT DEFLECTION TERMINALS:
  • Deflection sensitivity: 10 V/DIV or better
  • Input impedance: 2.2 MΩ shunted by 25 pF or less
  • POWER REQUIREMENTS: 117/230 VAC, 50/60 Hz, 16 watts
  • DIMENSIONS, OVERALL:
  • Width: 7-1/2" (190mm)
  • Height: 6" (154mm)
  • Depth: 11-27/32" (300mm)
  • WEIGHT: 8.36 lbs. (3.8 kg)

Safety & Warnings

Normal use of test equipment can expose users to electrical shock hazards, especially when testing near exposed voltage. An electrical shock with 10 milliamps of current passing through the heart can be fatal, and even voltages as low as 35 volts DC or AC RMS can be dangerous. To minimize risk, the following safety precautions should be observed:

1. Avoid unnecessary exposure to high voltage. Only remove housings and covers when necessary, and ensure equipment is turned off when making test connections in high voltage circuits.
2. Use insulated floor materials or large insulated mats, and ensure surfaces are dry. If unavailable, wear heavy gloves.
3. Employ the “one hand in the pocket” technique when handling instrument probes to avoid creating a ground path through the body.
4. Use an isolation transformer for transformerless “hot chassis” equipment to prevent shock hazards and potential damage to test instruments.
5. For equipment with a 3-wire AC power plug, use only a 3-wire outlet to maintain earth grounding.
6. Familiarize yourself with the equipment and its high voltage points, but be aware that high voltage may appear unexpectedly in defective equipment.
7. Remember that AC line voltage may be present on some power input circuit points even when equipment is off.
8. Never work alone; have someone nearby to provide aid if necessary. CPR training is recommended.

Download PDF (1.8 MB)