The manual for the Bk precision models covers the setup and operation of hand-held frequency counters, including features such as field strength measurement, automatic hold, and various controls like power, range, and calibration switches. It also provides troubleshooting tips, such as maintaining battery health and ensuring proper signal input to prevent random frequency displays.
Manual Overview
The BK Precision manual for Models 103 and 104 provides detailed guidance on troubleshooting, calibration, and usage of these hand-held frequency counters. Both models are designed for ease of use with features like a 10-digit LCD, low power consumption, and a NiCd battery pack. Troubleshooting focuses on ensuring correct switch settings for frequency ranges and verifying signal input to avoid random frequency displays, which can occur due to high receiver circuit gain. Calibration involves adjusting a trimmer capacitor accessible from the front panel to fine-tune the time base oscillator, enhancing accuracy when reading known frequencies. Error diagnostics include checking battery status and ensuring proper antenna use to prevent internal damage. The manual also advises on deep cycling NiCd batteries to maintain capacity and inspecting them annually for corrosion. The RF signal strength is displayed via a bargraph, and input sensitivity varies across frequency ranges, with maximum input specified to prevent damage. Proper antenna selection and reception distance from transmitters are crucial for optimal performance.
Quick Issues Covered
- Diagnose power-up failures with initial LCD segment test.
- Calibrate time base using front panel trimmer capacitor.
- Use schematics for circuit tracing and troubleshooting.
Overview
The product is a series of hand-held frequency counters, specifically Models 103 and 104, designed for ease of use and dependable performance. They are compact, pocket-sized test instruments that include features such as field strength measurement, a 10-digit liquid crystal display, low power consumption, and a bargraph to show RF signal strength. The manual covers various aspects including introduction, specifications, features, controls, hints and tips, input sensitivity, RF signal strength bargraph, and frequency display resolution.
Specifications
- Frequency range: 10 Hz – 3 GHz
- Weight: 250g
- Size: 100 mm high x 68 mm wide x 31 mm deep
- Impedance: 1 * dual purpose BNC Socket
- 50 Ohms input for range 1 MHz to 3 GHz
- 1 Meg Ohm input for range 10 Hz to 50 MHz
- Case: Stamped aluminum with black anodized finish
- Battery: Internal 4 x AA 600 mAH NiCd pack
- Power: 9 VDC 300 mA
- Timebase: Less than 1 PPM at room temperature
Controls & Indicators
Model 103 Controls:
1. Power Switch – Turns the counter on and initiates a 2-second test of all the LCD segments.
2. Range Switch – Switch to the 300 MHz position for frequencies between 1 MHz and 300 MHz and to the 3 GHz position for frequencies between 10 MHz and 3 GHz.
3. Hold Button – Holds the current display and stops the counter from counting.
4. Gate Button – Selects the gate or measurement time. A longer gate time allows counting for a longer period and results in higher accuracy.
5. Calibration – The calibration adjustment opening allows access to the trimmer capacitor for adjusting the time base oscillator.
Model 104 Controls:
1. Power Switch – Turns the counter on and initiates a 2-second test of all the LCD segments.
2. Amp Switch – Selects either the 1 Meg Ohm high impedance amplifier or 50 Ohm amplifier and prescale.
3. Range Switch – Switch to the 300 MHz position for frequencies between 1 MHz and 300 MHz and to the 3 GHz position for frequencies between 10 MHz and 3 GHz.
4. Lite Switch – Turns the LCD back light on and off.
5. Filter Switch – Turns the filter on and off.
6. Function Button – Selects the frequency or period with four settings, including automatic hold.
7. Hold Button – Holds the current display and stops the counter from counting.
8. Gate Button – Selects the gate or measurement time for higher accuracy.
9. Calibration – The calibration adjustment opening allows access to the trimmer capacitor for adjusting the time base oscillator.