The operation manual for the Rohde & Schwarz SRN RC Generator covers the setup process, including adjusting the mechanical instrument's zero point, setting the device to the correct mains voltage, and configuring the frequency and output voltage settings. It also provides troubleshooting guidance by detailing the characteristics of the generator, such as frequency range, output impedance, and error limits, to ensure proper operation and address any potential issues.
Manual Overview
The operation manual for the Rohde & Schwarz SRN RC Generator Type BN 4084 provides detailed instructions for technicians on troubleshooting, diagnostics, calibration, and adjustments. The generator operates within a frequency range of 2 to 20,000 Hz, divided into four bands, and features an output voltage range of 0 to 30V, adjustable via an overspannungsregler. For troubleshooting, technicians should verify the mechanical zero-point of the instrument when the device is off, adjusting as necessary using the slotted screw in the instrument housing. Calibration involves setting the device to the appropriate mains voltage, which requires disassembling the unit to access the voltage selector. Error diagnostics are facilitated by the built-in voltmeter, which displays either terminal voltage or open-circuit voltage depending on the output configuration. The manual emphasizes the importance of using the correct fuses and provides guidance on adjusting the output voltage and frequency settings. Technicians are advised to ensure proper load conditions to maintain accuracy, especially when using the internal capacitor to achieve DC-free output.
Quick Issues Covered
- Diagnose mechanischen Nullpunkt des Instruments bei ausgeschaltetem Gerät.
- Fehlergrenzen der Spannungsanzeige: +2% vom Endwert.
- Netzspannungseinstellung: 115, 125, 220, 235V mit Feinsicherung.
Overview
The RC-Generator Type SRN BN 4084 is a versatile device designed for a frequency range of 2 to 20,000 Hz, divided into four ranges. It features a logarithmic scale, unsymmetrical outputs, and a voltage range of 0 to 30V. It is equipped with a high voltage regulator and a chain divider for precise voltage output. The generator is suitable for various applications, including testing frequency responses and generating periodic stimulus currents for physiological studies. The manual covers the device’s features, applications, setup, and operation instructions, including adjusting mechanical instrument zero points, setting network voltage, frequency, and output voltage.
Specifications
- Frequenzbereich 2… 20000 Hz
- unterteilt in 4 Bereiche: 2…20/200/2000/20000 Hz
- Skalenverlauf etwa logarithmisch
- Fehlergrenzen: +2%
- Ausgänge unsymmetrisch, 2 Rändelklemmen und 1 koaxiale 13-mm-Buchse parallelgeschaltet
- Spannungsbereich 0…30V bzw. +30 db, stetig regelbar
- Innenwiderstand 0…5kΩ, je nach Regelstellung
- Außenwiderstand = 10 kΩ
- Spannungsbereich 10µV…3V bzw. -100…+12 db
- Innenwiderstand 600 Ω
- Fehlergrenzen des Teilers: +3% +10 µV
- Anzeige der Ausgangsspannung: durch Instrument mit den Bereichen 0…3V, 0…10V und -20…+2 db
- Fehlergrenzen der Spannungsanzeige: +2% vom Endwert
- Frequenzgang der Ausgangsspannung: +2%
- Klirrfaktor: <1% von 2…10Hz, <0,5% von 10…20000 Hz
- Brummspannung: <0,1% vom Bereichsnennwert
- Netzanschluss: 115/125/220/235V + 10%; 47…63 Hz; etwa 45 VA
- Bestückung: Röhren EF 86, Röhre EL 84, Röhre EL 86, Stabilisator 85 A 2, Zwergglimmlampe 220 V
- Schmelzeinsatz 0,3 C DIN 41571 (für 220/235 V)
- Abmessungen: 470 x 275 x 260 mm (R&S-Normkasten Größe 46)
- Gewicht: etwa 18 kg
Controls & Indicators
The source does not provide specific information about the front-panel controls/indicators.
Installation / Setup
The setup and installation highlights for the RC-Generator Type SRN BN 4084 include:
1. Mechanical Instrument Zero Point Adjustment: Ensure the instrument’s pointer is on the zero line of the volt scales when the device is off. Adjust by turning the slotted screw in the instrument housing if necessary.
2. Setting and Switching On the Device: The device is preset for 220V. To adjust for 115, 125, or 235V, loosen the screws on the front panel, remove the device from its housing, and bridge the appropriate spring pair on the voltage selector with a fine fuse. Use a 0.3 C DIN 41571 fuse for 220/235V and a 0.6 C DIN 41571 fuse for 115/125V. Reinstall the device, connect to the power supply, and switch on using the toggle switch on the front panel. A glow lamp above the switch indicates the power status.
3. Frequency Setting: The device has one scale for all four frequency ranges (2…20 Hz, 20…200 Hz, 200…2000 Hz, and 2000…20,000 Hz). For frequencies in the higher ranges, multiply the scale value by 10, 100, or 1000 as needed.
4. Output Voltage Setting: Outputs are available through two terminal sockets and a parallel 13-mm coaxial socket. Use 4-mm plugs for terminal sockets and a 13-mm plug (Order No. FS 413/11) for the coaxial socket.
5. Voltage Regulation: In the 30 V/+30 db stage, outputs are connected to the high voltage regulator, and the instrument displays the terminal voltage. The internal resistance varies with the regulator setting, increasing to about 5 kΩ at maximum voltage. A 2-uF capacitor can be switched in to block DC voltage if required.
6. Voltage Divider Stages: In stages from 3 V/+10 db to 0.1 mV/-80 db, a 10-stage 600-Ω voltage divider is used, and the voltmeter measures the open-circuit voltage. This configuration is indicated by green labeling on the front panel.
Operation
The operation and basic usage steps of the RC-Generator Type SRN BN 4084 include:
1. Checking the mechanical instrument’s zero point: Ensure the pointer is at the zero mark on the volt scales when the device is off. Adjust if necessary using the slotted screw in the instrument housing.
2. Setting the mains voltage and turning on the device: The device is factory-set for 220 V. To adjust for 115, 125, or 235 V, loosen the screws on the front panel, remove the device from its housing, and adjust the voltage selector. Use the appropriate fuse for the selected voltage. Reassemble, connect to the mains, and switch on using the toggle switch on the front panel. A glow lamp above the switch indicates the power status.
3. Setting the frequency: The device has four frequency ranges (2…20 Hz, 20…200 Hz, 200…2000 Hz, 2000…20,000 Hz) with a single calibrated scale. Multiply the scale value by 10, 100, or 1000 for the higher ranges. The frequency stability is maintained despite voltage fluctuations or temperature changes.
4. Setting the output voltage: Outputs are available via two terminal sockets and a parallel coaxial socket. The 30 V/+30 db stage connects the outputs to a high-voltage regulator, with the instrument showing the terminal voltage. The internal resistance varies with the regulator setting. A toggle switch can introduce a 2-uF capacitor to block DC voltage from the outputs. For stages from 3 V/+10 db to 0.1 mV/-80 db, a 10-stage 600-ohm divider is used, with the voltmeter measuring the divider’s high voltage.