The service manual for the Rohde & Schwarz SMIQ 6 GHz extension covers the testing, repair, and adjustment of the E6GHZ board, which extends the frequency range to 3300-6400 MHz, and includes detailed procedures for setting up and calibrating various components such as mixers, filters, and amplifiers. It also provides troubleshooting guidance, recommending the use of a test program for initial error diagnosis and checking the serial interface for correct data transmission before conducting more detailed troubleshooting in the RF paths.
Manual Overview
The service manual for the Rohde & Schwarz SMIQ 6 GHz extension focuses on the testing, troubleshooting, and calibration of the E6GHz board, which extends the frequency range of the SMIQ03 to 3300-6400 MHz. Troubleshooting involves using a test program from the service kit for initial error diagnosis, checking data transmission via the serial interface, and examining RF paths. The manual provides detailed procedures for testing and adjusting various components, including reference voltages, amplifier operating points, and switching voltages for RFSWITCH 1 and 2. Calibration involves adjusting the RF level linearity, drain current in the power unit, and tunable filters. Diagnostic voltages are used as indicators due to their large tolerances. The manual specifies using instruments like DC voltmeters, RF power meters, and spectrum analyzers for measurements. Board-level guidance includes descriptions of signal paths, mixer operations, and filter bank configurations, with emphasis on maintaining specified signal levels and conditions for optimal performance.
Quick Issues Covered
- Diagnose spectral purity issues with IQOUT6 output signal.
- Test and adjust RF level linearity using trimmers R394 and R173.
- Use circuit diagrams for tracing and troubleshooting RF path faults.
Overview
The document provides service information for the E6 GHz Board by Rohde & Schwarz, specifically focusing on testing and repairing the board. It covers function descriptions, measuring instruments, troubleshooting, testing, adjustment, disassembly, assembly, and external interfaces. The E6 GHz board extends the frequency range of the SMIQ03 to 3300-6400 MHz, involving processes like signal conversion and filtering. The document also details the components and procedures for ensuring proper operation and calibration of the board.