Rohde & Schwarz SME Service Manual Book4 Chapt5 (PDF)

The Rohde & Schwarz SME service manual provides detailed instructions for setting up, operating, and troubleshooting the LF-Generator SM-B2, including components like the LF-Synthesizer, DTMF-Generator, and electronic level adjustment. It also includes diagnostic procedures and fault-finding techniques using various measurement devices to ensure proper signal output and functionality.

  • Brand: Rohde & Schwarz
  • Type: service manual
  • File size: 3.5 MB

Manual Overview

The service manual for the Rohde & Schwarz SME, specifically for the LF-Generator SM-B2 option, provides detailed instructions for troubleshooting, diagnostics, calibration, and adjustments at the board level. The LF-Generator module generates signals from 0.1 Hz to 500 kHz, utilizing components such as the LF-Synthesizer, DTMF-Generator, and electronic level adjustment. Troubleshooting involves using the diagnostic function to monitor various voltages via a multiplexer, with specific diagnostic points designated for dual installations. Error codes and potential issues are addressed through a systematic approach, checking components like the DC-Offset, clock generator, and amplifiers. Calibration procedures include verifying data transmission, adjusting amplitudes, and ensuring frequency accuracy. The manual provides guidance on setting the Serbus address and variant, with specific jumper configurations for different slots. Calibration steps involve precise adjustments using voltmeters to ensure output signals meet specified tolerances. The manual emphasizes the importance of using appropriate measurement tools, such as frequency counters and audio analyzers, to ensure accurate diagnostics and adjustments.

Quick Issues Covered

  • Diagnose issues with LF-Generator using multiplexer D16.
  • Check DC-Offset at INT2 and LF-Buchse with diagnostics 1203/1303.
  • Verify DTMF signal issues by checking quartz frequency at B2 Pin1.

Overview

The product is the ROHDE & SCHWARZ LF-Generator SM-B2, which serves as an extended modulation source providing signals from 0.1 Hz to 500 kHz with a resolution of 0.1 Hz. The manual covers various aspects of the LF-Generator, including its functional description, components like the LF-Synthesizer, DTMF-Generator, electronic level adjustment, and diagnostic features. It also provides guidance on testing, troubleshooting, and calibration procedures for the device.

Specifications

  • LF-Synthesizer
  • DTMF-Generator
  • Elektronische Pegeleinstellung
  • Diagnose
  • Frequenzbereich: 0.1 Hz to 500 kHz
  • Auflösung: 0.1 Hz
  • DDS-Baustein: 32-Bit-Addierer, zwei 32-Bit-Inkrementregister, Sinus ROM
  • Taktfrequenz: 2.147484 MHz
  • Sinus ROM Adressen: 13-Bit
  • Kurvenform EPROM Adressen: 12-Bit
  • Kurvenformen: Sinus, Rechteck, Dreieck, Rauschen
  • Tiefpass Grenzfrequenz: 600 kHz
  • DTMF-Taktfrequenz: 3.579545 MHz
  • Pegeleinstellung: D/A-Wandler DAC8143
  • Amplitude: 0…4 V
  • Auflösung: 1 mV
  • Diagnosespannungsbereich: -5 V to +5 V

Troubleshooting & Error Codes

The manual provides troubleshooting guidance for the LF-Generator SM-B2. It includes a diagnostic function that can be used to check individual signals. The A/D converter in the diagnostic function measures DC voltages, so the signal frequency should be low enough (e.g., 0.1 Hz) for the converter to follow the signal. The manual lists possible error sources for various issues:

1. No signal at INT2 (X50.A7) and LF socket:
– Check DC offset using diagnosis 1203/1303.
– Check clock generator using diagnosis 1201/1301.
– Check supply voltages using diagnosis 1204/1304, 1205/1305, 1206/1306, 1207/1307.

2. Faulty output signal at X50.A7:
– Check INT2 amplifier using diagnosis 1202/1302.
– Check DDS/DTMF switching.

3. Faulty signal at LF socket:
– Check LFOUT amplifier using diagnosis 1203/1303.
– Check electronic level adjustment.
– Check INT1/INT2 switching.

4. Faulty SIN signal:
– Check SIN signal after the filter.
– Check Sample & Hold circuit.

5. Faulty waveform:
– Check signal at D/A converter.
– Check control signals and EPROM.

6. Faulty DTMF signal:
– Check quartz frequency at B2 Pin1.
– Check DTMF amplifier.
– Check DDS/DTMF switching.

7. Faulty LFGEN1 at LF socket:
– Check INT1/INT2 switching.