Rohde & Schwarz FSEK11 Operation Manual (PDF)

The manual provides instructions for setting up and operating the FSE-K11 application firmware module, including procedures for loading data sets, configuring test setups, and performing various measurements such as phase/frequency error and carrier power. It also covers troubleshooting steps for errors and failures during measurements and how to access and edit limit lines within the GSM software.

  • Brand: Rohde & Schwarz
  • Model: FSEK11
  • Type: operation manual
  • File size: 1.4 MB

Manual Overview

The operation manual for the FSE-K11 module outlines procedures for conducting GSM/DCS/PCS base station tests. It emphasizes the importance of using internal limit lines derived from applicable standards, which are available as datasets for user access and modification. The manual provides a step-by-step guide for loading these datasets onto the analyzer, which involves copying the desired datasets from a CD to the disk or analyzer, and using the RECALL function to load them into the instrument’s limit line directory. Troubleshooting is addressed through a section on errors and failures during measurement, with specific instructions on aborting measurements either by user command or instrument error. Calibration and adjustments are facilitated through the selection and editing of limit lines, which can be accessed in both GSM software and analyzer mode. The manual also includes detailed instructions for setting up tests, measuring various parameters like phase/frequency error, carrier power, and spectrum analysis, ensuring accurate and reliable test results.

Quick Issues Covered

  • Diagnose phase/frequency errors in GSM base stations.
  • Identify spurious emissions in RX/TX bands.
  • Calibrate power vs time measurements for GSM standards.

Overview

The manual covers the application firmware module FSE-K11 for GSM/DCS/PCS base station tests, providing measurement capabilities according to transmission standards. It supplements the FSE-K10 and FSE-K11 manuals, detailing the use of internal limit lines derived from applicable standards, which are available as datasets for user access. The manual includes instructions for loading these datasets into the instrument and accessing them during various measurements, such as power vs time, modulation spectrum, transient spectrum, and spurious measurements. Additionally, it provides information on setting up tests, measuring various parameters, and general operation of the FSE-K11 option.

Installation / Setup

The source provides instructions on how to load data sets for limit lines used in the FSE-K10 and FSE-K11 options. To do this, users need to copy the desired data sets from a CD to a disk or analyzer, use the RECALL hardkey, enter the path via the EDIT PATH softkey, select “Lines” from the ITEMS TO RECALL menu, and choose the desired data set from the DATA SET LIST. The limit lines can then be accessed and edited through the GSM software or in the analyzer mode of the instrument.

Operation

To load the data sets for the FSE-K11, follow these steps:

1. Copy the desired data sets from the CD to the disk or analyzer.
2. Press the hardkey RECALL and enter the path (e.g., A:) using the softkey EDIT PATH.
3. Select the menu item “Lines” via the softkey ITEMS TO RECALL.
4. Select the desired data set using the softkey DATA SET LIST. This will copy all limit lines of the selected data sets to the limit line directory of the instrument.

The limit lines can be accessed and edited via the softkey EDIT in the right-hand side menus of the following measurements: Power vs time, Modulation spectrum, Transient spectrum, or Spurious. Outside GSM measurements, the limit lines are accessible in the analyzer mode of the instrument.

Download PDF (1.4 MB)