Installation / Setup
The installation/setup highlights for the 4100 Series Digital Ohmmeters include the following:
1. Line Voltage/Fuse Selection: Before operation, verify the instrument is set for the proper local AC line voltage and that the correct fuse is installed. For 105 to 125 VAC, use a 0.25 Amp Slo-blo fuse; for 210 to 250 VAC, use a 0.125 Amp Slo-blo fuse. Fuse values are doubled if any interfaces are installed.
2. Bench Use: The unit comes with necessary hardware for bench use, and no special instructions are needed. Familiarity with Sections 4, 5, and 6 of the manual is recommended before operation.
3. Rack Mounting: Optional brackets are available for mounting in a standard 19″ equipment rack. The unit must be supported on both sides along its entire length. Blank panels at least 1.75 inches high should be installed between units to ensure air flow. Ambient temperature should not exceed 50°C during operation or 70°C when power is removed.
4. Safety Precautions: Use a three-contact power plug with a continuous ground connection. Discontinuity in the ground lead may render the unit unsafe. Special precautions are required when testing inductive loads like transformers to avoid damage or injury.